Laser surface scanning for patient registration in intracranial image-guided surgery

被引:163
作者
Raabe, A [1 ]
Krishnan, R [1 ]
Wolff, R [1 ]
Hermann, E [1 ]
Zimmermann, M [1 ]
Seifert, V [1 ]
机构
[1] Univ Frankfurt, Dept Neurosurg, Neuroctr, D-60528 Frankfurt, Germany
关键词
image-guided surgery; neuronavigation; surface registration;
D O I
10.1097/00006123-200204000-00021
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To report our clinical experience with a new laser scanning-based technique of surface registration. We performed a prospective study to measure the calculated registration error and the application accuracy of laser surface registration for intracranial image-guided surgery in the clinical setting. METHODS: Thirty-four consecutive patients with different intracranial diseases were scheduled for intracranial image-guided surgery by use of a passive infrared surgical navigation system. Surface registration was performed by use of a Class I laser device that emits a visible laser beam. The Polaris camera system (Northern Digital, Waterloo, ON, Canada) detects the skin reflections of the laser, which the software uses to generate a virtual three-dimensional matrix of the anatomy of each patient. An advanced surface-matching algorithm then matches this virtual three-dimensional matrix to the three-dimensional magnetic resonance therapy data set. Registration error as calculated by the computer was noted. Application accuracy was assessed by use of the localization error for three distant anatomic landmarks. RESULTS: Laser surface registration was successful in all patients. For the surgical field, application accuracy was 2.4 +/- 1.7 mm (range, 1-9 mm). Application accuracy was higher for the surgical field of frontally located lesions (mean, 3.8 +/- 0.8 mm; n = 13) as compared with temporal, parietal, occipital, and infratentorial lesions (mean, 2.8 +/- 2.1 mm; n = 21). CONCLUSION: Laser scanning for surface registration is an accurate, robust, and easy-to-use method of patient registration for image-guided surgery.
引用
收藏
页码:797 / 801
页数:5
相关论文
共 29 条
[1]
Alp MS, 1998, NEUROL RES, V20, P31
[2]
Barnett G H, 1995, J Image Guid Surg, V1, P46, DOI 10.1002/(SICI)1522-712X(1995)1:1<46::AID-IGS7>3.0.CO
[3]
2-M
[4]
USE OF A FRAMELESS, ARMLESS STEREOTAXIC WAND FOR BRAIN-TUMOR LOCALIZATION WITH 2-DIMENSIONAL AND 3-DIMENSIONAL NEUROIMAGING [J].
BARNETT, GH ;
KORMOS, DW ;
STEINER, CP ;
WEISENBERGER, J .
NEUROSURGERY, 1993, 33 (04) :674-678
[5]
A METHOD FOR REGISTRATION OF 3-D SHAPES [J].
BESL, PJ ;
MCKAY, ND .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1992, 14 (02) :239-256
[6]
Bucholz R, 2000, Stud Health Technol Inform, V70, P39
[7]
Clinical use of the optical digitizer for intracranial neuronavigation [J].
Germane, IM ;
Villalobos, H ;
Silvers, A ;
Post, KD .
NEUROSURGERY, 1999, 45 (02) :261-269
[8]
CLINICAL USE OF A FRAMELESS STEREOTAXIC ARM - RESULTS OF 325 CASES [J].
GOLFINOS, JG ;
FITZPATRICK, BC ;
SMITH, LR ;
SPETZLER, RF .
JOURNAL OF NEUROSURGERY, 1995, 83 (02) :197-205
[9]
BrainLab VectorVision neuronavigation system: Technology and clinical experiences in 131 cases [J].
Gumprecht, HK ;
Widenka, DC ;
Lumenta, CB .
NEUROSURGERY, 1999, 44 (01) :97-104
[10]
Helm P A, 1998, Comput Aided Surg, V3, P51, DOI 10.1002/(SICI)1097-0150(1998)3:2<51::AID-IGS1>3.0.CO