Secondary neutron dose during proton therapy using spot scanning

被引:237
作者
Schneider, U
Agosteo, S
Pedroni, E
Besserer, J
机构
[1] City Hosp Triemli, Dept Radiat Oncol & Nucl Med, Div Phys Med, Zurich, Switzerland
[2] Politecn Milan, Dipartimento Ingn Nucl, I-20133 Milan, Italy
[3] Paul Scherrer Inst, Dept Radiat Med, Villigen, Switzerland
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2002年 / 53卷 / 01期
关键词
proton therapy; dynamic scanning; Monte Carlo simulations; neutrons; secondary cancer;
D O I
10.1016/S0360-3016(01)02826-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: During proton radiotherapy, secondary neutrons are produced by nuclear interactions in the material in the beam line before and after entering the patient. The dose equivalent deposited by these neutrons is usually not considered in routine treatment planning. In this study, we estimated the neutron dose in patients from a spot scanning beam line by, performing measurements and Monte Carlo simulations. Methods and Materials: Measurements of the secondary neutron dose were performed during irradiation of a water phantom with 177-MeV protons using a Bonner sphere and CR39 etch detectors. Additionally, Monte Carlo simulations were performed using the FLUKA code. Results: A comparison of our measurements with measurements taken at a beam line using the scatter foil technique shows a dose advantage of at least 10 for the spot scanning technique. In the region of the Bragg peak, the neutron dose equivalent can reach for a medium-sized target volume approximately 1% of the treatment dose. Neutron doses expected in healthy tissues of the patient (in the not-treated volume) are for large and medium target volumes, approximately 0.004 Sv and 0.002 Sv per treatment Gy, respectively. Conclusions: We conclude from the measurements and simulations that the dose deposited by secondary neutrons during proton radiotherapy using the spot scanning technique can be neglected in the treatment region. In the healthy tissue, the dose coming from neutrons (0.002 Sv per treatment Gy) is approximately a factor of two larger than during photon treatment (0.001 Sv). These contributions to the integral dose from neutrons are still very low when compared to the dose sparing that can be achieved by using a proton beam instead of photons. (C) 2002 Elsevier Science Inc.
引用
收藏
页码:244 / 251
页数:8
相关论文
共 15 条
[1]   Secondary neutron and photon dose in proton therapy [J].
Agosteo, S ;
Birattari, C ;
Caravaggio, M ;
Silari, M ;
Tosi, G .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :293-305
[2]  
[Anonymous], 1991, 1990 recommendations of the international commission on radiological protection ICRP publication 60
[3]  
[Anonymous], KEK REPORT P
[4]   Secondary dose exposures during 200 MeV proton therapy [J].
Binns, PJ ;
Hough, JH .
RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) :441-444
[5]   BEAM DELIVERY SYSTEMS FOR CHARGED-PARTICLES [J].
BLATTMANN, H .
RADIATION AND ENVIRONMENTAL BIOPHYSICS, 1992, 31 (03) :219-231
[6]  
Fasso A., 1994, Proceedings of the Fourth International Conference on Calorimetry in High Energy Physics, P493
[7]  
FASSO A, 1997, TISRP9705 CERN, P158
[8]  
Ferrari A., 1998, P WORKSHOP NUCL REAC, P424
[9]  
Ferrari A, 1994, P MC93 INT C MONT CA, P277
[10]   A treatment planning inter-comparison of proton and intensity modulated photon radiotherapy [J].
Lomax, AJ ;
Bortfeld, T ;
Goitein, G ;
Debus, J ;
Dykstra, C ;
Tercier, PA ;
Coucke, PA ;
Mirimanoff, RO .
RADIOTHERAPY AND ONCOLOGY, 1999, 51 (03) :257-271