Quantification of anions and cations in environmental water samples -: Measurements with capillary electrophoresis and indirect-UV detection

被引:46
作者
Hiissa, T
Sirén, H
Kotiaho, T
Snellman, M
Hautojärvi, A
机构
[1] VTT, Chem Technol, FIN-02044 Espoo, Finland
[2] Acad Finland, FIN-02044 Espoo, Finland
[3] Posiva Ltd, FIN-00100 Helsinki, Finland
关键词
water analysis; environmental analysis; inorganic anions; inorganic cations;
D O I
10.1016/S0021-9673(99)00508-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The aim of this study was to validate two separation methods for determination of inorganic anions and cations from natural waters with capillary electrophoresis (CE) by using indirect-UV detection. The research is related to method development for screening of groundwater samples obtained in site investigations for spent fuel of the Finnish nuclear industry. In CE analysis, anions were separated in pyromellitic acid (pH 7.7) in the order bromide, chloride, sulphate, nitrite, nitrate, fluoride and dihydrogenphosphate. Cations were separated at pH 3.6 after anions using an 18-crown-6-ether solution. In these analyses, ammonium migrated first followed by potassium, calcium, sodium and magnesium. The concentrations of the ions in the natural water samples were calculated by using two or three calibration curves made using reference solutions at concentration levels of 0.5-250 mg/l. The repeatabilities of the anion and cation methods were tested using laboratory-made reference sample mixtures with high and low salt concentrations. The limits of quantification in the analyses were between 0.02 and 0.1 mg/l, depending on the ion. Concentrations of ions tested in natural waters varied from a few milligrams to tens of grams per litre. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:403 / 411
页数:9
相关论文
共 5 条
[1]   Permanent coated capillaries with reversed electroosmotic flow for anion analysis [J].
Finkler, C ;
Charrel, H ;
Engelhardt, H .
JOURNAL OF CHROMATOGRAPHY A, 1998, 822 (01) :101-106
[2]  
HABER C, 1994, J CAPILLARY ELECTROP, V3, P313
[3]  
JANDIK P, 1993, CAPILLARY ELECTROPHO
[4]   Determination of low-molecular-mass anionic compounds in beverage samples using capillary zone electrophoresis with simultaneous indirect ultraviolet and conductivity detection [J].
Klampfl, CW ;
Katzmayr, MU .
JOURNAL OF CHROMATOGRAPHY A, 1998, 822 (01) :117-123
[5]   Determination of small anions by capillary electrophoresis using indirect UV detection with sulphonated nitrosonaphthol dyes [J].
Siren, H ;
Maattanen, A ;
Riekkola, ML .