An improved method for measuring vapor pressure by DSC with automated pressure control

被引:28
作者
Casserino, M [1 ]
Blevins, DR [1 ]
Sanders, RN [1 ]
机构
[1] ALBEMARLE CORP,BATON ROUGE,LA 70898
关键词
vapour pressure; DSC; thermal analysis;
D O I
10.1016/0040-6031(96)02923-1
中图分类号
O414.1 [热力学];
学科分类号
摘要
Vapor pressure measurements are necessary for a wide range of applications in the chemical industry. Traditional methods require 10-30 mi of sample, 6-8 h of operator time, and multiple measurements for high precision data. With a modification to the pressure differential scanning calorimeter cell, use of hermetic-type pans with pin-holed lids, and the addition of a pressure controller, vapor pressure measurements from 5-760 torr can be obtained. These measurements use less than 1 mi of sample, take about 2 h of operator time, and achieve high precision data on a single set of measurements.