In vivo measurement of radius calcium phosphorus ratio by X-ray absorptiometry

被引:46
作者
Fountos, G
Tzaphlidou, M
Kounadi, E
Glaros, D
机构
[1] Univ Ioannina, Sch Med, Med Phys Lab, GR-45110 Ioannina, Greece
[2] Univ Hosp Ioannina, Dept Med Phys, Ioannina 45110, Greece
关键词
D O I
10.1016/S0969-8043(99)00056-1
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
We describe a new method for assessing the skeletal Ca/P ratio in vivo using X-ray absorptiometry. By placing cerium (Ce) and samarium (Sm) filters in the X-ray beam from a commercial X-ray source (Norland), mean photon energies of 39 and 89 keV were obtained. The instrument was designed to take measurements of the forearm, at a site located at the distal 1/3 of the radius. The system was calibrated with three bone phantoms: Ca-10(PO4)(6)(OH)(2), Ca(HPO4)(H2O)(2) and Ca(HPO4)(2)(H2O)). The precision for measuring the Ca/P ratio in the human radius was 2.3% CV for a skin dose to the forearm ranging from 0.3 to 0.4 mGy, depending on the width of the arm. The Ca/P ratio of the radius was significantly lower in patients with postmenopausal osteoporosis than in premenopausal controls. (C0 1999 Elsevier Science Ltd. All rights reserved.