Thin scintillators and position sensitive photomultiplier tubes for hard X-ray imaging in space

被引:11
作者
Kroeger, RA [1 ]
Grove, JE [1 ]
Inderhees, SE [1 ]
Johnson, WN [1 ]
Kinzer, RL [1 ]
Kurfess, JD [1 ]
Phlips, BP [1 ]
机构
[1] UNIV SPACE RES ASSOC,WASHINGTON,DC
关键词
D O I
10.1109/23.603770
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Monolithic and segmented CsI(TI) scintillators coupled to a position sensitive photomultiplier tube (PSPMT) have been tested for imaging and environmental performance necessary for hard X-ray astrophysics missions. Using a crossed-wire PSPMT and individual anode-wire readout, spatial resolution of 1-2 mm rms at 60 keV over the entire photo-cathode imaging surface is obtained out to near the edges of the scintillator. A PSPMT mapping correction was used to obtain energy resolution close to what is expected from CsI(Tl) scintillators.
引用
收藏
页码:881 / 884
页数:4
相关论文
共 4 条
[1]   MULTICHANNEL READOUT OF CROSSED-WIRE ANODE PHOTOMULTIPLIERS [J].
BIRD, AJ ;
HE, Z ;
RAMSDEN, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3) :668-672
[2]  
HARRISON FA, 1990, SPIE, V1344, P47
[3]  
JOHNSON WN, 1996, SPIE, V2806, P20
[4]  
OHASHI T, 1991, P SOC PHOTO-OPT INS, V1549, P9, DOI 10.1117/12.48322