Limits to magnetic resonance microscopy

被引:131
作者
Glover, P [1 ]
Mansfield, P [1 ]
机构
[1] Univ Nottingham, Magnet Resonance Ctr, Nottingham NG7 2RD, England
关键词
D O I
10.1088/0034-4885/65/10/203
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The last quarter of the twentieth century saw the development of magnetic resonance imaging (MRI) grow from a laboratory demonstration to a multi-billion dollar worldwide industry. There is a clinical body scanner in almost every hospital of the developed nations. The field of magnetic resonance microscopy (MRM), after mostly being abandoned by researchers in the first decade of MRI, has become an established branch of the science. This paper reviews the development of MRM over the last decade with an emphasis on the current state of the art. The fundamental principles of imaging and signal detection are examined to determine the physical principles which limit the available resolution. The limits are discussed with reference to liquid, solid and gas phase microscopy. In each area, the novel approaches employed by researchers to push back the limits of resolution are discussed. Although the limits to resolution are well known, the developments and applications of MRM have not reached their limit.
引用
收藏
页码:1489 / 1511
页数:23
相关论文
共 66 条
  • [1] NUCLEAR-MAGNETIC-RESONANCE IMAGING OF A SINGLE CELL
    AGUAYO, JB
    BLACKBAND, SJ
    SCHOENIGER, J
    MATTINGLY, MA
    HINTERMANN, M
    [J]. NATURE, 1986, 322 (6075) : 190 - 191
  • [2] [Anonymous], 1990, PRINCIPLES MAGNETIC, DOI DOI 10.1007/978-3-662-09441
  • [3] Single-point ramped imaging with T-1 enhancement (SPRITE)
    Balcom, BJ
    MacGregor, RP
    Beyea, SD
    Green, DP
    Armstrong, RL
    Bremner, TW
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES A, 1996, 123 (01) : 131 - 134
  • [4] Magnetic resonance diffraction using the magnetic field from a ferromagnetic sphere
    Barbic, M
    [J]. JOURNAL OF APPLIED PHYSICS, 2002, 91 (12) : 9987 - 9994
  • [5] Fast high-resolution magnetic resonance imaging demonstrates fractality of myocardial perfusion in microscopic dimensions
    Bauer, WR
    Hiller, KH
    Galuppo, P
    Neubauer, S
    Köpke, J
    Haase, A
    Waller, C
    Ertl, G
    [J]. CIRCULATION RESEARCH, 2001, 88 (03) : 340 - 346
  • [6] BLUMLER P, 1998, SPATIALLY RESOLVED M
  • [7] Multilayer gradient coil design
    Bowtell, R
    Robyr, P
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1998, 131 (02) : 286 - 294
  • [8] RESOLUTION OF CELLULAR STRUCTURES BY NMR MICROSCOPY AT 11.7-T
    BOWTELL, RW
    BROWN, GD
    GLOVER, PM
    MCJURY, M
    MANSFIELD, P
    [J]. PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 1990, 333 (1632): : 457 - 467
  • [9] Callaghan PT., 1993, Principles of Nuclear Magnetic Resonance Microscopy
  • [10] Chen CN, 1989, BIOMEDICAL MAGNETIC