Isotype controls in phenotyping and quantification of microparticles: A major source of error and how to evade it

被引:28
作者
Trummer, Arne [1 ]
De Rop, Christiane [1 ]
Tiede, Andreas [1 ]
Ganser, Arnold [1 ]
Eisert, Roswith [1 ]
机构
[1] Hannover Med Sch, Dept Hematol Hemostasis & Oncol, D-30625 Hannover, Germany
关键词
microparticles; flow cytometry; isotype controls; P-selectin;
D O I
10.1016/j.thromres.2008.01.005
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The characterisation and quantification of cell-derived microparticles (MPs) using flow cytometry are often complicated by a tow staining intensity and a non-discrete signal pattern of many cell surface antigens. Fluorescence-labelled isotype controls (ICs) are commonly used to set limits for the discrimination of antigen positive vs. negative events. Objectives: The influence of different ICs on the characterisation and quantification of MPs was studied. Antigen negative MPs stained with an antibody of interest were evaluated as an alternative control. Methods: MPs were prepared from platelets, endothelial cell Lines and leucemic cell lines and stained with fluorescein isothiocyanate (FITC) or phycoerythrin (PE) Labelled antibodies or isotype controls. Results are given as the mean fluorescence intensity (MFI) or percentage of "false-positive" events above a fluorescence intensity >1. Results: Using identical instrument settings, seven different ICs (FITC-conjugates N = 3, PE-conjugatesN=4) resulted in a wide range of MFI and percentage of positive events with a mean coefficient of variation (CV) of 0.77. Instead, NMPs showed less variability with a mean CV of 0.50 and allowed a reliable and reproducible quantification of MPs when set as controls with <2% false-positive events above an FI >1. As a result, the expression of certain antigens (e.g. CD62P) was lower compared to previous reports in the literature. Conclusions: Diversity in the staining intensity of isotype controls is a potential source of error in the characterisation and quantification of MPs by flow cytometry. The use of antigen negative MPs to adjust instrument settings is suggested. (C) 2008 Elsevier Ltd. ALL rights reserved.
引用
收藏
页码:691 / 700
页数:10
相关论文
共 22 条
[1]   More on: Cellular microparticles: what are they bad or good for? [J].
Ahn, YS ;
Jy, W ;
Jimenez, JJ ;
Horstman, LL .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2004, 2 (07) :1215-1216
[2]   Proteome of endothelial cell-derived procoagulant microparticles [J].
Banfi, C ;
Brioschi, M ;
Wait, R ;
Begum, S ;
Gianazza, E ;
Pirillo, A ;
Mussoni, L ;
Tremoli, E .
PROTEOMICS, 2005, 5 (17) :4443-4455
[3]  
Baranska Patrycja, 2005, Cancer Genomics & Proteomics, V2, P265
[4]   High levels of circulating endothelial microparticles in patients with acute coronary syndromes [J].
Bernal-Mizrachi, L ;
Jy, W ;
Jimenez, JJ ;
Pastor, J ;
Mauro, LM ;
Horstman, LL ;
de Marchena, E ;
Ahn, YS .
AMERICAN HEART JOURNAL, 2003, 145 (06) :962-970
[5]   The adhesion molecule P-selectin and cardiovascular disease [J].
Blann, AD ;
Nadar, SK ;
Lip, GYH .
EUROPEAN HEART JOURNAL, 2003, 24 (24) :2166-2179
[6]  
Bode AP, 2000, PLATELETS, V11, P259
[7]   A new role in hemostasis for the adhesion receptor P-selectin [J].
Cambien, B ;
Wagner, DD .
TRENDS IN MOLECULAR MEDICINE, 2004, 10 (04) :179-186
[8]   MUC18, a member of the immunoglobulin superfamily, is expressed on bone marrow fibroblasts and a subset of hematological malignancies [J].
Filshie, RJA ;
Zannettino, ACW ;
Makrynikola, V ;
Gronthos, S ;
Henniker, AJ ;
Bendall, LJ ;
Gottlieb, DJ ;
Simmons, PJ ;
Bradstock, KF .
LEUKEMIA, 1998, 12 (03) :414-421
[9]   Role of platelet P-selectin and microparticle PSGL-1 in thrombus formation [J].
Furie, B ;
Furie, BC .
TRENDS IN MOLECULAR MEDICINE, 2004, 10 (04) :171-178
[10]   The platelet microparticle proteome [J].
Garcia, BA ;
Smalley, DM ;
Cho, HJ ;
Shabanowitz, J ;
Ley, M ;
Hunt, DF .
JOURNAL OF PROTEOME RESEARCH, 2005, 4 (05) :1516-1521