The peri-operative use of the oesophageal Doppler monitor in patients undergoing coronary artery revascularisation - A comparison with the continuous cardiac output monitor

被引:15
作者
Krishnamurthy, B [1 ]
McMurray, TJ [1 ]
McClean, E [1 ]
机构
[1] ROYAL GRP HOSP, DEPT CLIN ANAESTHESIA, BELFAST BT12 6BA, ANTRIM, NORTH IRELAND
关键词
measurement technique; oesophageal Doppler; surgery; cardiac;
D O I
10.1111/j.1365-2044.1997.147-az0151.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
This study was carried out to evaluate the ease of use and reliability of cardiac output estimations performed by an oesophageal Doppler monitor and to compare its use with that of a continuous cardiac output pulmonary flotation catheter. Measurements were made during and after surgery in 16 patients scheduled to undergo coronary revascularisation. Both devices suffered significant intra-operative problems which led us to question their suitability as operating theatre monitors. After surgery the continuous cardiac output monitor provided stable results while the oesophageal Doppler monitor required the continuous presence of an experienced anaesthetist to ensure comparable cardiac output estimations.
引用
收藏
页码:624 / 629
页数:6
相关论文
共 13 条
  • [1] BARASH PG, 1981, J THORAC CARDIOV SUR, V82, P5
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] The effectiveness of right heart catheterization in the initial care of critically ill patients
    Connors, AF
    Speroff, T
    Dawson, NV
    Thomas, C
    Harrell, FE
    Wagner, D
    Desbiens, N
    Goldman, L
    Wu, AW
    Califf, RM
    Fulkerson, WJ
    Vidaillet, H
    Broste, S
    Bellamy, P
    Lynn, J
    Knaus, WA
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11): : 889 - 897
  • [4] COMPLICATIONS OF PULMONARY-ARTERY CATHETERIZATION IN THE CARE OF CRITICALLY ILL PATIENTS - PROSPECTIVE-STUDY
    ELLIOTT, CG
    ZIMMERMAN, GA
    CLEMMER, TP
    [J]. CHEST, 1979, 76 (06) : 647 - 652
  • [5] MARK JB, 1986, ANESTH ANALG, V65, P1013
  • [6] MYTHEN MG, 1995, ARCH SURG-CHICAGO, V130, P423
  • [7] PAYEN D, 1994, INT P J, V1, P3
  • [8] REICH DL, 1993, CARDIAC ANESTHESIA, P261
  • [9] NONINVASIVE OPTIMIZATION OF LEFT-VENTRICULAR FILLING USING ESOPHAGEAL DOPPLER
    SINGER, M
    BENNETT, ED
    [J]. CRITICAL CARE MEDICINE, 1991, 19 (09) : 1132 - 1137
  • [10] SINGER M, 1994, INT P J, V1, P1