Application of a hexapole collision and reaction cell in ICP-MS Part I: Instrumental aspects and operational optimization

被引:146
作者
Feldmann, I [1 ]
Jakubowski, N [1 ]
Stuewer, D [1 ]
机构
[1] Inst Spektrochem & Angew Spekt, D-44013 Dortmund, Germany
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1999年 / 365卷 / 05期
关键词
D O I
10.1007/s002160051633
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The application of an ion-guiding buffer gas-filled hexapole collision and reaction cell in ICP-MS has been studied in order to give a preliminary performance characterization of a new instrument providing this feature for increasing the ion yield and decreasing contributions from Ar induced interfering molecular ions. As buffer gas He was used while H-2 served as reaction gas. Addition of the latter can be an effective means for reduction of typical argon induced polyatomic ions (Ar+, ArO+, Ar-2(+)) by orders of magnitude owing to gas phase reactions. Molecular interferences generated in the cell can be suppressed by a retarding electric field established by a de hexapole bias potential of -2 V.
引用
收藏
页码:415 / 421
页数:7
相关论文
共 18 条