Use of a micro-ionization chamber and an anthropomorphic head phantom in a quality assurance program for stereotactic radiosurgery

被引:16
作者
Duggan, DM
Coffey, CW
机构
[1] Center for Radiation Oncology, Vanderbilt Clinic, Vanderbilt University Medical Center, Nashville
关键词
stereotactic radiosurgery; quality assurance; micro-ionization techniques;
D O I
10.1118/1.597682
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Quality assurance methods used in association with radiosurgery must include all aspects of the radiosurgery process: visualization and localization of the target, treatment and dose planning and dose delivery. Presented here is a quality assurance method that utilizes an anthromorphic head phantom and a micro-ionization chamber to demonstrate precise target localization and accurate dose delivery. This micro-ionization chamber method offers an immediate readout which is both accurate and reproducible. Additionally, this method allows unlimited repetition of the dose measurement process without repeated radiographic localization studies as is necessary with the conventional methods of TLD, film, and Fricke gels. The method and techniques presented can be used in the acceptance testing and routine quality assurance of both linac-based and Gamma Knife radiosurgery units. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:513 / 516
页数:4
相关论文
共 8 条
  • [1] CURRENT COLLECTION AND IONIC RECOMBINATION IN SMALL CYLINDRICAL IONIZATION CHAMBERS EXPOSED TO PULSED RADIATION
    BOAG, JW
    CURRANT, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (629) : 471 - 478
  • [2] COFFEY CW, 1993, STEREOT FUNCT NEUROS, V61, P130, DOI 10.1159/000100667
  • [3] ASSURANCE OF HIGH-QUALITY LINAC-BASED STEREOTAXIC RADIOSURGERY
    DRZYMALA, RE
    KLEIN, EE
    SIMPSON, JR
    RICH, KM
    WASSERMAN, TH
    PURDY, JA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (02): : 459 - 472
  • [4] DUGGAN DM, UNPUB MED PHYS
  • [5] CYLINDRICAL CHAMBER DIMENSIONS AND THE CORRESPONDING VALUES OF AWALL AND NGAS/(NXAION)
    GASTORF, R
    HUMPHRIES, L
    ROZENFELD, M
    [J]. MEDICAL PHYSICS, 1986, 13 (05) : 751 - 754
  • [6] CALCULATED RESPONSE AND WALL CORRECTION FACTORS FOR IONIZATION CHAMBERS EXPOSED TO CO-60 GAMMA-RAYS
    NATH, R
    SCHULZ, RJ
    [J]. MEDICAL PHYSICS, 1981, 8 (01) : 85 - 93
  • [7] ASSESSMENT OF THE ACCURACY OF STEREOTAXIC RADIOSURGERY USING FRICKE-INFUSED GELS AND MRI
    SCHULZ, RJ
    MARYANSKI, MJ
    IBBOTT, GS
    BOND, JE
    [J]. MEDICAL PHYSICS, 1993, 20 (06) : 1731 - 1734
  • [8] 1983, MED PHYS, V10, P1