Pacemaker and ICD generator malfunctions - Analysis of food and drug administration annual reports

被引:228
作者
Maisel, WH
Moynahan, M
Zuckerman, BD
Gross, TP
Tovar, OH
Tillman, DB
Schultz, DB
机构
[1] Harvard Univ, Sch Med, Beth Israel Deaconess Med Ctr, Div Cardiovasc, Boston, MA 02215 USA
[2] US FDA, Ctr Devices & Radiol Hlth, Rockville, MD 20857 USA
来源
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION | 2006年 / 295卷 / 16期
关键词
D O I
10.1001/jama.295.16.1901
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Context Pacemakers and implantable cardioverter-defibrillators (ICDs) are complex medical devices proven to reduce mortality in specific high-risk patient populations. It is not known if increasing device complexity is associated with decreased reliability. Objectives To analyze postapproval annual reports submitted to the US Food and Drug Administration (FDA) by manufacturers of pacemakers and ICDs to determine the reported number and rate of pacemaker and ICD malfunctions and to assess trends in device performance. Design and Setting Pacemaker and ICD annual reports submitted to the FDA for the years 1990-2002 were reviewed. A pacemaker or ICD generator was defined as having malfunctioned if it was explanted due to an observed malfunction, returned to the manufacturer, and confirmed by the manufacturer to be functioning inappropriately. Leads and biventricular devices were not included in the study. Deaths were attributed to device malfunction only if they were witnessed, the malfunction immediately led to the death, and the malfunction was confirmed by the manufacturer. Main Outcome Measures Number of implanted pacemaker and ICD generators; number of reported malfunctions; and annual malfunction replacement rates. Generator malfunction replacement rates were defined as the annual number of replacements due to confirmed malfunction divided by the annual number of implants. Results During the study period, 2.25 million pacemakers and 415780 ICDs were implanted in the United States. Overall, 17 323 devices (8834 pacemakers and 8489 ICDs) were explanted due to confirmed malfunction. Battery/capacitor abnormalities (4085 malfunctions [23.6%]) and electrical issues (4708 malfunctions [27.1%]) accounted for half of the total device failures. The annual pacemaker malfunction replacement rate per 1000 implants decreased significantly during the study, from a peak of 9.0 in 1993 to a low of 1.4 in 2002 (P=.006 for trend). In contrast, the ICD malfunction replacement rate per 1000 implants, after decreasing from 38.6 in 1993 to 7.9 in 1996, increased markedly during the latter half of the study, peaking in 2001 at 36.4 (P=.04 for trend). More than half of the reported ICD malfunctions occurred in the last 3 years of the study. Overall, the annual ICD malfunction replacement rate was significantly higher than the pacemaker malfunction replacement rate (mean [SD], 20.7 [11.6] vs 4.6 [2.2] replacements per 1000 implants; P<.001; rate ratio, 5.9 [95% confidence interval, 2.7-9.1]). Sixty-one deaths (30 pacemaker patients, 31 ICD patients) were attributable to device malfunction. Conclusions This study demonstrates that thousands of patients have been affected by pacemaker and ICD malfunctions, the pacemaker malfunction replacement rate has decreased, the ICD malfunction replacement rate increased during the latter half of the study, and the ICD malfunction replacement rate is significantly higher than that for pacemakers. Although pacemakers and ICDs are important life-sustaining devices that have saved many lives, careful monitoring of device performance is still required.
引用
收藏
页码:1901 / 1906
页数:6
相关论文
共 21 条
  • [1] Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure
    Bardy, GH
    Lee, KL
    Mark, DB
    Poole, JE
    Packer, DL
    Boineau, R
    Domanski, M
    Troutman, C
    Anderson, J
    Johnson, G
    McNulty, SE
    Clapp-Channing, N
    Davidson-Ray, LD
    Fraulo, ES
    Fishbein, DP
    Luceri, RM
    Ip, JH
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) : 225 - 237
  • [2] Electrophysiologic testing to identify patients with coronary artery disease who are at risk for sudden death
    Buxton, AE
    Lee, KL
    DiCarlo, L
    Gold, MR
    Greer, GS
    Prystowsky, EN
    O'Toole, MF
    Tang, A
    Fisher, JD
    Coromilas, J
    Talajic, M
    Hafley, G
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (26) : 1937 - 1945
  • [3] Safety alerts involving device therapy for arrhythmias
    Eagle, KA
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2001, 286 (07): : 843 - 844
  • [4] Gregoratos G, 2002, ACC AHA NASPE 2002 G
  • [5] *GUID CORP, 2004, GUID CARD RHYTHM MAN
  • [6] Feasibility and initial results of an Internet-based pacemaker and ICD pulse generator and lead registry
    Hauser, R
    Hayes, D
    Parsonnet, V
    Furman, S
    Epstein, A
    Hayes, J
    Saksena, S
    Irwin, M
    Almquist, A
    Cannom, D
    Gross, J
    Kallinen, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (01): : 82 - 87
  • [7] Kutalek SP, 2000, CLIN CARDIAC PACING, P645
  • [8] Pacemaker and ICD generator reliability - Meta-analysis of device registries
    Maisel, WH
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (16): : 1929 - 1934
  • [9] Safety issues involving medical devices - Implications of recent implantable cardioverter-defibrillator malfunctions
    Maisel, WH
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 294 (08): : 955 - 958
  • [10] Medical device regulation: An introduction for the practicing physician
    Maisel, WH
    [J]. ANNALS OF INTERNAL MEDICINE, 2004, 140 (04) : 296 - 302