Noninvasive ventilator triggering in chronic obstructive pulmonary disease - A test lung comparison

被引:41
作者
Stell, IM
Paul, G
Lee, KC
Ponte, J
Moxham, J
机构
[1] Kings Coll London Hosp, Guys Kings & St Thomas Sch Med, Dept Accid & Emergency Med, London, England
[2] Kings Coll London Hosp, Guys Kings & St Thomas Sch Med, Dept Anaesthet, London, England
[3] Kings Coll London Hosp, Guys Kings & St Thomas Sch Med, Dept Resp Med, London, England
[4] Kings Coll London, Dept Mech Engn, London WC2R 2LS, England
关键词
intermittent positive-pressure ventilation; obstructive lung diseases; patient acceptance of health care;
D O I
10.1164/ajrccm.164.11.2011147
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
To be most effective, noninvasive ventilation (NIV) ventilators should synchronize well with patients' breathing. However, the speed with which different ventilators can respond to the transitions between inspiration and expiration may vary, and abnormal respiratory mechanics and mask leaks may exacerbate this problem. This study explored synchronization using a new test lung model designed to simulate acute exacerbations of chronic obstructive pulmonary disease (COPD). Thirteen ventilators were tested against different combinations of tidal volume (VT), airways resistance (Raw), FRC, and mask leak. These combinations ranged from those of a severe exacerbation of COPD, to a mild condition reflecting the optimal triggering conditions a ventilator is likely to encounter. The triggering delays from the beginning and end of "inspiration" of the test lung, to the appropriate responses from the ventilators were measured. Three of the ventilators had trigger delays less than approximately 120 ms at both the beginning and end of expiration under all conditions. Trigger delays of other ventilators were mainly in the range of 120 to 300 ms, although exceptionally as long as 500 ms. Varying the conditions had a variable but generally small effect on triggering times, suggesting that there is a largely unavoidable element to the triggering delays intrinsic to the design of the ventilators.
引用
收藏
页码:2092 / 2097
页数:6
相关论文
共 25 条
[1]  
Adams A B, 2000, Respir Care, V45, P390
[2]   Partitioning of inspiratory muscle workload and pressure assistance in ventilator-dependent COPD patients [J].
Appendini, L ;
Purro, A ;
Patessio, A ;
Zanaboni, S ;
Carone, M ;
Spada, E ;
Donner, CF ;
Rossi, A .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 154 (05) :1301-1309
[3]   NONINVASIVE VENTILATION FOR ACUTE EXACERBATIONS OF CHRONIC OBSTRUCTIVE PULMONARY-DISEASE [J].
BROCHARD, L ;
MANCEBO, J ;
WYSOCKI, M ;
LOFASO, F ;
CONTI, G ;
RAUSS, A ;
SIMONNEAU, G ;
BENITO, S ;
GASPARETTO, A ;
LEMAIRE, F ;
ISABEY, D ;
HARF, A .
NEW ENGLAND JOURNAL OF MEDICINE, 1995, 333 (13) :817-822
[4]   Performance characteristics of bilevel pressure ventilators - A lung model study [J].
Bunburaphong, T ;
Imanaka, H ;
Nishimura, M ;
Hess, D ;
Kacmarek, RM .
CHEST, 1997, 111 (04) :1050-1060
[5]   Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger [J].
Calderini, E ;
Confalonieri, M ;
Puccio, PG ;
Francavilla, N ;
Stella, L ;
Gregoretti, C .
INTENSIVE CARE MEDICINE, 1999, 25 (07) :662-667
[6]  
ELLIOTT MW, 1993, ANAESTHESIA, V48, P8
[7]  
FERGUSON GT, 1995, AM J RESP CRIT CARE, V151, P1126
[8]  
HILL NS, 1996, RESP CARE, V41, P885
[9]   Pathophysiologic basis of acute respiratory distress in patients who fail a trial of weaning from mechanical ventilation [J].
Jubran, A ;
Tobin, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1997, 155 (03) :906-915
[10]   VARIABILITY OF PATIENT-VENTILATOR INTERACTION WITH PRESSURE SUPPORT VENTILATION IN PATIENTS WITH CHRONIC OBSTRUCTIVE PULMONARY-DISEASE [J].
JUBRAN, A ;
VANDEGRAAFF, WB ;
TOBIN, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 152 (01) :129-136