Calibration and application of an X-ray image intensifier/charge-coupled device detector for monochromatic macromolecular crystallography

被引:59
作者
Hammersley, AP [1 ]
Brown, K [1 ]
Burmeister, W [1 ]
Claustre, L [1 ]
Gonzalez, A [1 ]
McSweeney, S [1 ]
Mitchell, E [1 ]
Moy, JP [1 ]
Svensson, SO [1 ]
Thompson, AW [1 ]
机构
[1] EMBL OUTSTN, F-38042 GRENOBLE, FRANCE
关键词
detector calibration; data collection; detector distortions; X-ray detectors; charge-coupled device detectors; macromolecular crystallography;
D O I
10.1107/S0909049596015087
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Charge-coupled device (CCD)-based X-ray detectors allow data to be collected much more quickly (similar to 10 times) than with current on-line imaging-plate systems. At the ESRF, X-ray image intensifier/CCD detector systems have been developed. These have great potential as fast read-out detectors for macromolecular and other forms of crystallography. They are relatively large sensitive X-ray detectors but have two inherent weaknesses: convex detection surfaces leading to spatial distortion and non-uniformity of intensity response, and susceptibility to small changes in magnetic fields. A large improvement has been made to the accuracy obtained by non-uniformity of response calibration and correction, using fluorescence from doped lithium borate glasses. Monochromatic macromolecular crystallography demonstration experiments with external user groups have shown that high-quality results may be obtained under real experimental conditions.
引用
收藏
页码:67 / 77
页数:11
相关论文
共 33 条
[1]   THE CCP4 SUITE - PROGRAMS FOR PROTEIN CRYSTALLOGRAPHY [J].
BAILEY, S .
ACTA CRYSTALLOGRAPHICA SECTION D-BIOLOGICAL CRYSTALLOGRAPHY, 1994, 50 :760-763
[2]  
BRUNGER AT, 1993, X PLOR SYSTEM CRYSTA
[3]   FORMAT ALTERATIONS IN CCD BASED ELECTROOPTIC X-RAY-DETECTORS [J].
DECKMAN, HW ;
GRUNER, SM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3) :527-533
[4]  
FANCHON E, 1995, ESRF NEWSLETT, V24, P6
[5]  
FOURME R, 1991, P EUR WORKSH XRAY DE
[6]  
FOUST S, 1995, EXPSF199501 ESRF
[7]  
HAEDENER A, 1997, UNPUB ACTA CRYST D
[8]  
Hammersley A. P., 1995, EXPAH9501 ESRF
[9]  
HAMMERSLEY AP, 1995, REV SCI INSTRUM, V66, P2729, DOI [10.1063/1.1145618, 10.1063/1.1146453]
[10]   CALIBRATION AND CORRECTION OF SPATIAL DISTORTIONS IN 2D DETECTOR SYSTEMS [J].
HAMMERSLEY, AP ;
SVENSSON, SO ;
THOMPSON, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 346 (1-2) :312-321