Silicon-on-insulator microfluidic device with monolithic sensor integration for μTAS applications

被引:13
作者
Sharma, S [1 ]
Buchholz, K
Luber, SM
Rant, U
Tornow, M
Abstreiter, G
机构
[1] Univ London Imperial Coll Sci Technol & Med, Inst Biomed Engn, London SW7 2AZ, England
[2] Tech Univ Munich, Walter Schottky Inst, D-85748 Garching, Germany
关键词
field-effect transistor (FET); lab-on-chip (LOC); microfluidics; micrototal analysis system (mu TAS); silicon-on-insulator (SOI);
D O I
10.1109/JMEMS.2006.872222
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel concept for the integration of liquid phase charge sensors into microfluidic devices based on silicon-on-insulator (SOI) technology is reported. Utilizing standard silicon processing we fabricated basic microfluidic cross geometries comprising of 5-10-mm-long and 55-mu m-wide channels of 3 mu m. depth by wet sacrificial etching of the buried oxide of an SOI substrate. To demonstrate the feasibility of fluid manipulation along the channel we performed electroosmotic pumping of a dye-labeled buffer solution. At selected positions along the channel we patterned the 205-nm thin top silicon layer into freely suspended, 10-mu m wide bars bridging the channel. We demonstrate how these monolithically integrated bars work as thin-film resistors that sensitively probe changes of the surface potential via the field effect. In this way, a combination of electrokinetic manipulation and separation of charged analytes together with an on-chip electronic detection can provide a new basis for the label-free analysis of, e.g., biomolecular species as envisaged in the concept of micrototal analysis systems (mu TAS) or Lab-on-Chip (LOC).
引用
收藏
页码:308 / 313
页数:6
相关论文
共 22 条
[1]  
Alarie JP, 2000, ELECTROPHORESIS, V21, P100, DOI 10.1002/(SICI)1522-2683(20000101)21:1<100::AID-ELPS100>3.3.CO
[2]  
2-G
[3]   Micro total analysis systems. 2. Analytical standard operations and applications [J].
Auroux, PA ;
Iossifidis, D ;
Reyes, DR ;
Manz, A .
ANALYTICAL CHEMISTRY, 2002, 74 (12) :2637-2652
[4]  
Becker H, 2000, ELECTROPHORESIS, V21, P12, DOI 10.1002/(SICI)1522-2683(20000101)21:1<12::AID-ELPS12>3.3.CO
[5]  
2-Z
[6]   Thirty years of ISFETOLOGY - What happened in the past 30 years and what may happen in the next 30 years [J].
Bergveld, P .
SENSORS AND ACTUATORS B-CHEMICAL, 2003, 88 (01) :1-20
[7]  
Bergveld P., 1988, ANAL CHEM, VXXIII
[8]   Modular concept of a laboratory on a chip for chemical and biochemical analysis [J].
Blankenstein, G ;
Larsen, UD .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (3-4) :427-438
[9]   Micromachining of buried micro channels in silicon [J].
de Boer, MJ ;
Tjerkstra, RW ;
Berenschot, JW ;
Jansen, HV ;
Burger, CJ ;
Gardeniers, JGE ;
Elwenspoek, M ;
van den Berg, A .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2000, 9 (01) :94-103
[10]   Integrated microfluidic devices [J].
Erickson, D ;
Li, DQ .
ANALYTICA CHIMICA ACTA, 2004, 507 (01) :11-26