Gaps in the Traceability Chain of Human Growth Hormone Measurements

被引:18
作者
Boulo, Sebastien [1 ]
Hanisch, Katja [1 ]
Bidlingmaier, Martin [2 ]
Arsene, Cristian-Gabriel [3 ]
Panteghini, Mauro [4 ]
Auclair, Guy [1 ]
Sturgeon, Catharine [5 ]
Schimmel, Heinz [1 ]
Zegers, Ingrid [1 ]
机构
[1] European Commiss, Joint Res Ctr, IRMM, B-2440 Geel, Belgium
[2] Klinikum Univ Munchen, Med Klin & Poliklin 4, Endocrine Lab, Munich, Germany
[3] PTB, Braunschweig, Germany
[4] Univ Milan, Ctr Metrol Traceabil Lab Med, Milan, Italy
[5] Royal Infirm, UK Natl External Qual Assessment Serv NEQAS, Dept Lab Med, Edinburgh, Midlothian, Scotland
关键词
CONSENSUS STATEMENT; STANDARDIZATION; ASSAY; QUANTIFICATION; INTERFERENCE; SOMATROPIN; SERUM; GH;
D O I
10.1373/clinchem.2012.199489
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Human growth hormone (hGH) is measured for the diagnosis of secretion disorders. These measurements fall under the EU Directive 98/79/EC on in vitro diagnostic medical devices requiring traceability of commercial calibrator values to higher-order reference materials or procedures (Off J Eur Communities 1998 Dec 7; L 331:1-37). External quality assessment schemes show large discrepancies between results from different methods, even though most methods provide results traceable to the recommended International Standard (IS 98/574). The aim of this study was to investigate possible causes for these discrepancies. METHODS: We investigated the commutability and recovery of hGH in reconstituted IS 98/574. We tested different reconstitution protocols and used 4 different serum matrices for spiking. These IS preparations were measured together with serum samples. We quantified hGH by 5 different methods in 4 different laboratories. RESULTS: Results from the different methods correlated well for the serum samples. Mean discrepancies between results from different methods were <= 20%. None of the IS preparations was commutable for all the method comparisons. The recovery of hGH in preparations of IS 98/574 depended on the reconstitution protocol (>10-fold differences) and background matrix (relative differences <= 17% for different serum matrices). CONCLUSIONS: The use of different protocols for reconstitution and spiking of hGH reference preparations affects quantification by immunoassays, potentially leading to a bias between commercial methods, despite the use of calibrators with values claimed to be traceable to the same higher-order reference material. (C) 2013 American Association for Clinical Chemistry
引用
收藏
页码:1074 / 1082
页数:9
相关论文
共 23 条
[1]  
[Anonymous], CERTIFICATION PROTEI
[2]  
[Anonymous], 2012, R LANG ENV STAT COMP
[3]  
[Anonymous], 2011, D405211 ASTM
[4]   Quantification of growth hormone in serum by isotope dilution mass spectrometry [J].
Arsene, Cristian G. ;
Henrion, Andre ;
Diekmann, Nina ;
Manolopoulou, Jenny ;
Bidlingmaier, Martin .
ANALYTICAL BIOCHEMISTRY, 2010, 401 (02) :228-235
[5]   Biochemical assessment and long-term monitoring in patients with acromegaly: Statement from a joint consensus conference of The Growth Hormone Research Society and The Pituitary Society [J].
Bengtsson, BA ;
Carlsson, L ;
Christianssen, JS ;
Clemmons, D ;
Frohman, L ;
Ho, K ;
Mullis, PE ;
Robinson, I ;
Strasburger, CJ ;
Thorner, MO ;
Vance, ML ;
von Werder, K ;
Freda, PU ;
Holdaway, I ;
Johansson, G ;
Jorgensen, JOL ;
Wass, J ;
Barkan, A ;
Brue, TC ;
Chihara, K ;
Colao, A ;
de Herder, WW ;
Malozowski, S ;
Molitch, ME ;
Orskov, H ;
Schopohl, J ;
Shalet, SM ;
Stewart, PM ;
Swearingen, B ;
Trainer, PJ ;
Anderberg, B ;
Baister, L ;
Friend, KE ;
Harris, PE ;
Hartman, ML ;
Ispas-Jouron, S ;
Kappelgaard, AM ;
Knudsen, KW ;
Konno, T ;
Loumaye, E ;
Musgrave, D ;
Olivier, C ;
Pedroncelli, A ;
Shimatsu, A .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2004, 89 (07) :3099-3102
[6]   Measurement of human growth hormone by immunoassays: Current status, unsolved problems and clinical consequences [J].
Bidlingmaier, Martin ;
Freda, Pamela U. .
GROWTH HORMONE & IGF RESEARCH, 2010, 20 (01) :19-25
[7]   The second international standard for somatropin (recombinant DNA-derived human growth hormone): Preparation and calibration in an international collaborative study [J].
Bristow, AF ;
Jespersen, AM .
BIOLOGICALS, 2001, 29 (02) :97-106
[8]   Consensus Statement on the Standardization and Evaluation of Growth Hormone and Insulin-like Growth Factor Assays [J].
Clemmons, David R. .
CLINICAL CHEMISTRY, 2011, 57 (04) :555-559
[9]  
CLSI, 2010, C53A CLSI
[10]   Growth hormone isoforrns and segments/fragments: Molecular structure and laboratory measurement [J].
De Palo, EF ;
De Filippis, V ;
Gatti, R ;
Spinella, P .
CLINICA CHIMICA ACTA, 2006, 364 (1-2) :67-76