Linear hydraulic pressure-pulse actuator (LHPA): A versatile instrument that produces a simulated blood pressure pulse wave for small sized vessels

被引:1
作者
Field, S [1 ]
Drzewiecki, G [1 ]
机构
[1] RUTGERS STATE UNIV,DEPT BIOMED ENGN,CARDIOVASC DYNAM LAB,PISCATAWAY,NJ 08855
关键词
D O I
10.1109/10.495288
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An instrument is presented which produces a simulated circulatory pulsatile pressure wave for small sized vessels. The linear hydraulic pressure-pulse actuator (LHPA) is designed to be extremely versatile, that is, a blood pressure wave source of any shape, amplitude, offset and frequency can be simulated. In addition, the LHPA can reproduce accurately a real pulse pressure wave by simply inputting an actual data record of a circulatory pressure pulse. The design is accomplished by incorporating the use of a linear force solenoid driven with a voltage-to-current source power amplifier. Testing of the device is presented here, as well as pressure pulse results from a recorded pulsatile pressure input to the LHPA. The device is simple to implement in that its response is linear, for volume changes upto +/-5 mL, without the need for feedback compensation.
引用
收藏
页码:663 / 668
页数:6
相关论文
共 8 条
[1]   EFFECT OF STENOSIS ON WALL MOTION - A POSSIBLE MECHANISM OF STROKE AND TRANSIENT ISCHEMIC ATTACK [J].
BINNS, RL ;
KU, DN .
ARTERIOSCLEROSIS, 1989, 9 (06) :842-847
[2]   PRESSURE-FLOW RELATIONSHIPS IN COLLAPSIBLE TUBES [J].
CONRAD, WA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1969, BM16 (04) :284-&
[3]   ARTERIAL TONOMETRY - REVIEW AND ANALYSIS [J].
DRZEWIECKI, GM ;
MELBIN, J ;
NOORDERGRAAF, A .
JOURNAL OF BIOMECHANICS, 1983, 16 (02) :141-152
[4]   FLUID-DYNAMICS OF CORONARY-ARTERY STENOSIS [J].
MATES, RE ;
GUPTA, RL ;
BELL, AC ;
KLOCKE, FJ .
CIRCULATION RESEARCH, 1978, 42 (01) :152-162
[5]   A FINITE-ELEMENT SIMULATION OF PULSATILE FLOW IN FLEXIBLE OBSTRUCTED TUBES [J].
ROOZ, E ;
YOUNG, DF ;
ROGGE, TR .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1982, 104 (02) :119-124
[6]   ALTERED CORONARY FLOW RESPONSES TO VASOACTIVE DRUGS IN THE PRESENCE OF CORONARY ARTERIAL-STENOSIS IN THE DOG [J].
SANTAMORE, WP ;
WALINSKY, P .
AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02) :276-285
[7]  
SIEBES M, 1989, P IEEE ENG MED BIOL, P101
[8]   FLOW CHARACTERISTICS IN MODELS OF ARTERIAL STENOSES .2. UNSTEADY FLOW [J].
YOUNG, DF ;
TSAI, FY .
JOURNAL OF BIOMECHANICS, 1973, 6 (05) :547-559