Biomechanical Evaluation of an Interspinous Stabilizing Device, Locker

被引:180
作者
Shim, Chan Shik [1 ]
Park, Seoung Woo [2 ]
Lee, Sang-Ho [1 ]
Lim, T. Jesse [3 ]
Chun, Kwonsoo [3 ]
Kim, Daniel H. [4 ]
机构
[1] Wooridul Spine Hosp, Dept Neurosurg, Seoul 135100, South Korea
[2] Kangwon Natl Univ, Coll Med, Dept Neurosurg, Chunchon, South Korea
[3] Stanford Univ, Med Ctr, Dept Surg, Stanford, CA 94305 USA
[4] Baylor Coll Med, Dept Surg, Houston, TX 77030 USA
关键词
biomechanics; posterior dynamic stabilization; spine arthroplasty; interspinous Locker;
D O I
10.1097/BRS.0b013e3181894fb1
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A biomechanical study. Objective. To evaluate the biomechanical properties of a novel dynamic interspinous device named Locker. Summary of Background Data. There has been no biomechanical study on this device. Methods. Five human cadaveric lumbar spine specimens (L2-S1) were tested in the following sequence: (1) intact state; (2) after removal of the interspinous ligaments of L3-L4; (3) after application of the Locker at the interspinous space; (4) after destabilization of the L3-L4 motion segments; and (5) after reapplication of the Locker to the destabilized segments. Range of motion (ROM) and intradiscal pressure were measured by a video-based motion capture system and needle transducers. Results. Applied to the intact spine, the Locker significantly reduced the ROM of the segment in extension and flexion without significant effect on lateral bending and/or axial rotation. When applied to the destabilized segment, it significantly reduced the ROM in all directions of movement except in axial rotation, where it failed to restore the ROM of the segment. The ROM of adjacent segments was not significantly affected by the application of the Locker. Pressures at the posterior anulus and central nucleus were decreased by application of the Locker. Conclusion. The Locker showed a significant stabilizing effect on the spinal motion segment both in the intact and destabilized spine without any significant effect on adjacent segments. It also decreased the pressures of the posterior anulus and central nucleus significantly. However, in the destabilized spine, it showed no stabilizing effect in axial rotation. This biomechanical property should be considered in clinical application.
引用
收藏
页码:E820 / E827
页数:8
相关论文
共 32 条
[1]
Elastic stabilization alone or combined with rigid fusion in spinal surgery: a biomechanical study and clinical experience based on 82 cases [J].
Caserta, S ;
La Maida, GA ;
Misaggi, B ;
Peroni, D ;
Pietrabissa, R ;
Raimondi, MT ;
Redaelli, A .
EUROPEAN SPINE JOURNAL, 2002, 11 (Suppl 2) :S192-S197
[2]
Dynamic interspinous process technology [J].
Christie, SD ;
Song, JK ;
Fessler, RG .
SPINE, 2005, 30 (16) :S73-S78
[3]
The effect of spinal destabilization and instrumentation on lumbar intradiscal pressure an in vitro biomechanical analysis [J].
Cunningham, BW ;
Kotani, Y ;
McNulty, PS ;
Cappuccino, A ;
McAfee, PC .
SPINE, 1997, 22 (22) :2655-2663
[4]
DUBOIS G, 2006, DYNAMIC RECONSTRUCTI, P299
[5]
Development and preclinical testing of a new tension-band device for the spine: the Loop system [J].
Garner, MD ;
Wolfe, SJ ;
Kuslich, SD .
EUROPEAN SPINE JOURNAL, 2002, 11 (Suppl 2) :S186-S191
[6]
MECHANICAL-PROPERTIES OF LUMBAR SPINAL MOTION SEGMENTS AS AFFECTED BY PARTIAL DISK REMOVAL [J].
GOEL, VK ;
NISHIYAMA, K ;
WEINSTEIN, JN ;
LIU, YK .
SPINE, 1986, 11 (10) :1008-1012
[7]
GUIGUI P, 1994, REV CHIR ORTHOP, V80, P681
[8]
INSTANTANEOUS AXIS OF ROTATION AS A FUNCTION OF THE 3 COLUMNS OF THE SPINE [J].
HAHER, TR ;
OBRIEN, M ;
FELMLY, WT ;
WELIN, D ;
PERRIER, G ;
CHOUEKA, J ;
DEVLIN, V ;
VASSILIOU, A ;
CHOW, G .
SPINE, 1992, 17 (06) :S149-S154
[9]
Kaech DL, 2002, SPINAL RESTABILIZATION PROCEDURES: DIAGNOSTIC AND THERAPEUTIC ASPECTS OF INTERVERTEBRAL FUSION CAGES, ARTIFICIAL DISCS AND MOBILE IMPLANTS, P355
[10]
LUMBOSACRAL SPINAL-FUSION - A BIOMECHANICAL STUDY [J].
LEE, CK ;
LANGRANA, NA .
SPINE, 1984, 9 (06) :574-581