Locking mechanism failure in semiconstrained total elbow arthroplasty

被引:21
作者
Figgie, MP [1 ]
Su, EP [1 ]
Kahn, B [1 ]
Lipman, J [1 ]
机构
[1] Hosp Special Surg, Comprehens Arthritis Serv, New York, NY 10021 USA
关键词
D O I
10.1016/j.jse.2005.05.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Between 1988 and 2000, we performed 170 total elbow replacements using a semiconstrained implant (custom or standard). We are aware of 9 implants (5.3%) that have had late failure of the locking mechanism, as evidenced by migration of the connecting axle on radiographic examination. The mean time to failure of the locking mechanism was 5.5 years. Eight elbows required surgery for revision of the polyethylene bushing and locking mechanism by use of a custom-made locking clip. A limited approach was devised that allowed sufficient exposure to change the bushing and axle. In 1 of the 8 cases, the revised locking mechanism failed at 5.5 years, necessitating a reoperation. Of the 8 patients, 7 have not had recurrence of the failure at a mean of 1.4 years (minimum follow-up of 1 year). These 7 patients have good or excellent results based on the Hospital for Special Surgery elbow scoring system, with a mean arc of motion of 130 degrees.
引用
收藏
页码:88 / 93
页数:6
相关论文
共 13 条
[1]
BRYAN RS, 1982, CLIN ORTHOP RELAT R, P188
[2]
Total elbow arthroplasty in patients who have juvenile rheumatoid arthritis [J].
Connor, PM ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (05) :678-688
[3]
Figgie H E 3rd, 1986, J Arthroplasty, V1, P169, DOI 10.1016/S0883-5403(86)80027-4
[4]
FIGGIE HE, 1987, CLIN ORTHOP RELAT R, P185
[5]
The Coonrad-Morrey total elbow arthroplasty in patients who have rheumatoid arthritis - A ten to fifteen-year follow-up study [J].
Gill, DRJ ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (09) :1327-1335
[6]
TOTAL ELBOW REPLACEMENT [J].
INGLIS, AE ;
PELLICCI, PM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (08) :1252-1258
[7]
THE PRITCHARD MARK-II ELBOW PROSTHESIS IN RHEUMATOID-ARTHRITIS [J].
MADSEN, F ;
GUDMUNDSON, GH ;
SOJBJERG, JO ;
SNEPPEN, O .
ACTA ORTHOPAEDICA SCANDINAVICA, 1989, 60 (03) :249-253
[8]
Matarese W, 1990, J Arthroplasty, V5, P365, DOI 10.1016/S0883-5403(08)80096-4
[9]
SEMICONSTRAINED ARTHROPLASTY FOR THE TREATMENT OF RHEUMATOID-ARTHRITIS OF THE ELBOW [J].
MORREY, BF ;
ADAMS, RA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (04) :479-490
[10]
SEMICONSTRAINED ELBOW REPLACEMENT FOR DISTAL HUMERAL NONUNION [J].
MORREY, BF ;
ADAMS, RA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (01) :67-72