Mechanical performance of acrylic bone cements containing different radiopacifying agents

被引:120
作者
Ginebra, MP
Albuixech, L
Fernández-Barragán, E
Aparicio, C
Gil, FJ
San Román, J
Vázquez, B
Planell, JA
机构
[1] Univ Politecn Catalunya, Dept Mat Sci & Met Engn, Biomed Engn Res Ctr, Barcelona 08028, Spain
[2] CSIC, Inst Ciencia & Tecnol Polimeros, E-28006 Madrid, Spain
关键词
acrylic bone cements; PMMA; radiopacifying agents; mechanical properties; fatigue crack propagation;
D O I
10.1016/S0142-9612(01)00314-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effect that three different radiopacifying agents, two of them inorganic (BaSO4. ZrO2) and one organic (an iodine containing monomer, 1HQM) have on the static and dynamic mechanical properties of acrylic bone cements was studied. Compressive and tensile strength, fracture toughness and fatigue crack propagation were evaluated. The effect of the inorganic fillers depends on their size and morphology. In relation to the radiolucent cement, the addition of zirconium dioxide improved significantly the tensile strength, the fracture toughness and the fatigue crack propagation resistance. In contrast, the addition of barium sulphate produced a decrease of the tensile strength, but did not affect the fracture toughness and improved the crack propagation resistance. When the iodine containing monomer was used, although the tensile strength and the fracture toughness increased, the fatigue crack propagation resistance remained as low as it was for the radiolucent cement. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1873 / 1882
页数:10
相关论文
共 50 条
[1]  
American Society for Testing and Materials (ASTM), 1996, F45195 ASTM, P49
[2]  
[Anonymous], 1989, ANN BOOK ASTM STAND, P487
[3]  
*ASTM, 1989, ANN BOOK ASTM STAND, P646
[4]  
Beaumont P W, 1979, J Biomed Eng, V1, P147, DOI 10.1016/0141-5425(79)90035-9
[5]   STUDIES ON A NEW RADIOPAQUE POLYMERIC BIOMATERIAL [J].
BENZINA, A ;
KRUFT, MAB ;
BAR, F ;
VANDERVEEN, FH ;
BASTIAANSEN, CW ;
HEIJNEN, V ;
REUTELINGSPERGER, C ;
KOOLE, LH .
BIOMATERIALS, 1994, 15 (14) :1122-1128
[6]   MICROMECHANISMS OF FATIGUE-CRACK INITIATION AND PROPAGATION IN BONE CEMENTS [J].
BHAMBRI, SK ;
GILBERTSON, LN .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (02) :233-237
[7]  
Caravia L, 1990, Proc Inst Mech Eng H, V204, P65, DOI 10.1243/PIME_PROC_1990_204_230_02
[8]   CERAMIC BEARING SURFACES IN TOTAL ARTIFICIAL JOINTS - RESISTANCE TO 3RD BODY WEAR DAMAGE FROM BONE-CEMENT PARTICLES [J].
COOPER, JR ;
DOWSON, D ;
FISHER, J ;
JOBBINS, B .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1991, 15 (02) :63-67
[9]  
Davy KWM, 1997, POLYM INT, V43, P143, DOI 10.1002/(SICI)1097-0126(199706)43:2<143::AID-PI717>3.0.CO
[10]  
2-M