An automatic method for identifying the location of the papillary muscles in two-dimensional (2-D) short-axis echocardiographic images is described, The technique uses both spatial and temporal information to identify the presence and track the location of the muscles in the left ventricle from end-diastole to end-systole, The three main steps of the method are spatial preprocessing, spatial processing, and temporal processing, The spatial preprocessing step includes a region of search estimation, The spatial processing step includes a papillary muscle existence test and an initial approximation of the papillary muscle points, The temporal processing includes motion-pattern evaluation and final papillary muscle location, The estimates of existence and position for the automatic method were compared with estimates made by an independent expert observer, Two hundred and ten frames, three taken from each of 70 image sequences, were evaluated, Since two regions of search were processed for each frame (one for the posterior-inferior and one for the anterior-lateral papillary muscle), a total of 420 approximations were made, Of this total, 340 automatic estimates were judged to be in close agreement with estimates made by the expert, Of the remaining 80 approximations, 54 estimates were made by the expert when the computer determined that no papillary muscle was present, 17 estimates provided poor results, and nine estimates were made by the computer when the observer concluded that no papillary muscle was present.
引用
收藏
页码:460 / 470
页数:11
相关论文
共 22 条
[1]   SEMIAUTOMATED BORDER TRACKING OF CINE ECHOCARDIOGRAPHIC VENTRICULAR IMAGES [J].
ADAM, D ;
HAREUVENI, O ;
SIDEMAN, S .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1987, 6 (03) :266-271
[2]   CONTRAST ENHANCEMENT TECHNIQUE BASED ON LOCAL DETECTION OF EDGES [J].
BEGHDADI, A ;
LENEGRATE, A .
COMPUTER VISION GRAPHICS AND IMAGE PROCESSING, 1989, 46 (02) :162-174
[3]  
BOSCH JG, 1988, P 15 C COMP CARD, P137
[4]   DETECTING LEFT-VENTRICULAR ENDOCARDIAL AND EPICARDIAL BOUNDARIES BY DIGITAL TWO-DIMENSIONAL ECHOCARDIOGRAPHY [J].
CHU, CH ;
DELP, EJ ;
BUDA, AJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (02) :81-90
[5]  
Duncan J. S., 1991, Proceedings 1991 IEEE Computer Society Conference on Computer Vision and Pattern Recognition (91CH2983-5), P318, DOI 10.1109/CVPR.1991.139709
[6]   AUTOMATIC VENTRICULAR CAVITY BOUNDARY DETECTION FROM SEQUENTIAL ULTRASOUND IMAGES USING SIMULATED ANNEALING [J].
FRIEDLAND, N ;
ADAM, D .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (04) :344-353
[7]  
Geiser E A, 1990, J Am Soc Echocardiogr, V3, P79
[8]  
GEISER EA, 1986, ECHOCARDIOGRAPHY PHY
[9]  
Gonzalez RC, 1987, Digital Image Processing, V2nd
[10]   FEATURE ENHANCEMENT OF FILM MAMMOGRAMS USING FIXED AND ADAPTIVE NEIGHBORHOODS [J].
GORDON, R ;
RANGAYYAN, RM .
APPLIED OPTICS, 1984, 23 (04) :560-564