Reconstruction method for fluorescent X-ray computed tomography by least-squares method using singular value decomposition

被引:101
作者
Yuasa, T
Akiba, M
Takeda, T
Kazama, M
Hoshino, A
Watanabe, Y
Hyodo, K
Dilmanian, FA
Akatsuka, T
Itai, Y
机构
[1] UNIV TSUKUBA,INST CLIN MED,TSUKUBA,IBARAKI 305,JAPAN
[2] NATL LAB HIGH ENERGY PHYS,PHOTON FACTORY,TSUKUBA,IBARAKI 305,JAPAN
[3] BROOKHAVEN NATL LAB,DEPT MED,UPTON,NY 11973
关键词
computed tomography; fluorescent X-ray; least-squares method; Monte Carlo simulation; singular value decomposition; synchrotron radiation;
D O I
10.1109/23.554824
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We describe a new attenuation correction method for fluorescent X-ray computed tomography (FXCT) applied to image nonradioactive contrast materials in vivo. The principle of the FXCT imaging is that of computed tomography of the first generation, Using monochromatized synchrotron radiation from the BLNE-5A bending-magnet beam line of Tristan Accumulation Ring in KEK, Japan, we studied phantoms with the FXCT method, and we succeeded in delineating a 4-mm-diameter channel filled with a 500 mu g I/ml iodine solution in a 20-mm-diameter acrylic cylindrical phantom, However, to detect smaller iodine concentrations, attenuation correction is needed, We present a correction method based on the equation representing the measurement process, The discretized equation system is solved by the least-squares method using the singular value decomposition, The attenuation correction method is applied to the projections by the Monte Carlo simulation and the experiment to confirm its effectiveness.
引用
收藏
页码:54 / 62
页数:9
相关论文
共 16 条
[1]   FLUORESCENCE TOMOGRAPHY USING SYNCHROTRON RADIATION AT THE NSLS [J].
BOISSEAU, P ;
GRODZINS, L .
HYPERFINE INTERACTIONS, 1987, 33 (1-4) :283-292
[2]  
BRANHAM RL, 1990, SCI DATA ANAL INTRO, P34
[3]   A NEW TOMOGRAPHIC DEVICE BASED ON THE DETECTION OF FLUORESCENT X-RAYS [J].
CESAREO, R ;
MASCARENHAS, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 277 (2-3) :669-672
[4]   THE VALIDITY OF MONTE-CARLO SIMULATION IN STUDIES OF SCATTERED RADIATION IN DIAGNOSTIC-RADIOLOGY [J].
CHAN, HP ;
DOI, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (02) :109-129
[5]  
Groetsch C. W., 1993, INVERSE PROBLEMS MAT, P84
[6]   FLUORESCENT COMPUTER-TOMOGRAPHY - A MODEL FOR CORRECTION OF X-RAY ABSORPTION [J].
HOGAN, JP ;
GONSALVES, RA ;
KRIEGER, AS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (06) :1721-1727
[7]  
Hubbell J. H., 1975, Journal of Physical and Chemical Reference Data, V4, P471, DOI 10.1063/1.555523
[8]  
Iida A., 1991, HDB SYNCHROTRON RAD, V4, P307
[9]   METHODS OF LEAST-SQUARES AND SIRT IN RECONSTRUCTION [J].
LAKSHMINARAYANAN, AV ;
LENT, A .
JOURNAL OF THEORETICAL BIOLOGY, 1979, 76 (03) :267-295
[10]  
LAWSON CL, 1993, SIAM, P107