Prime Time for Enzymatic Creatinine Methods in Pediatrics

被引:86
作者
Cobbaert, Christa M. [1 ]
Baadenhuijsen, Henk [2 ]
Weykamp, Cas W. [3 ]
机构
[1] Amphia Hosp, Dept Clin Chem & Hematol, NL-4800 RK Breda, Netherlands
[2] UMC St Radbound, SKML, Nijmegen, Netherlands
[3] Queen Beatrix Hosp, Dept Clin Chem, Winterswijk, Netherlands
关键词
SERUM CREATININE; JAFFE REACTION; WORKING GROUP; INTERFERENCE; BILIRUBIN; ASSAYS; IMPROVEMENT; DISEASE;
D O I
10.1373/clinchem.2008.116863
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
100118 [医学信息学]; 100208 [临床检验诊断学];
摘要
BACKGROUND: The literature regarding the nonspecificity of applications of the Jaffe (alkaline picrate) reaction for creatinine is generally outdated. We conducted a specificity study to update the nonspecificity information for current Jaffe and enzymatic creatinine assays. METHODS: Two serum pools with creatinine concentrations within the pediatric reference interval were spiked with albumin, IgG, unconjugated bilirubin, adult hemoglobin (Hb A), and fetal hemoglobin (Hb F) to produce 1 unspiked and 5 spiked samples per pool. The 35 laboratories participating in the survey used a total of 14 method-analyzer combinations. Measurements were performed in triplicate in a single run in accordance with manufacturer instructions. Absolute differences in creatinine concentration between spiked and unspiked samples were calculated per laboratory. Mixed ANOVA was used to quantify the interferent-related CV component for the Jaffe and enzymatic methods. RESULTS: The interference by bilirubin and Hb A on serum creatinine measurements was < 10% for most of the Jaffe and enzymatic methods. Obvious interference was observed among the Jaffe methods in samples spiked with Hb F, albumin, and IgG, but not among the enzymatic methods. The within-laboratory interferent-related CVs for the Jaffe method-analyzer combinations ranged from 8.0%-27% at a creatinine concentration of 40.4 mu mol/L (0.46 mg/dL) and from 5.4%-15% at 73.4 mu mol/L (0.83 mg/dL). Enzymatic methods had within-laboratory interferent-related CVs of < 4% at both concentrations. CONCLUSIONS: Albumin, IgG, and Hb F interfered with Jaffe creatinine assays, leading to inaccuracies in estimated glomerular filtration rates that are clinically important, especially in children and neonates. Because protein error and Hb F interference do not occur with any of the enzymatic methods tested, we conclude that enzymatic creatinine methods are preferred for evaluation of kidney function in pediatric cases. (c) 2008 American Association for Clinical Chemistry
引用
收藏
页码:549 / 558
页数:10
相关论文
共 23 条
[1]
[Anonymous], 1997, LINEAR MIXED MODELS, DOI DOI 10.1007/978-1-4612-2294-1_4
[2]
BOWERS LD, 1980, CLIN CHEM, V26, P555
[3]
BOYNE P, 1985, CLIN CHEM, V31, P1564
[4]
BOYNE P, 1988, CLIN BIOCHEM REV, V9, P54
[5]
Reference intervals for serum creatinine concentrations: Assessment of available data for global application [J].
Ceriotti, Ferruccio ;
Boyd, James C. ;
Klein, Gerhard ;
Henny, Joseph ;
Queralto, Josep ;
Kairisto, Veli ;
Panteghini, Mauro .
CLINICAL CHEMISTRY, 2008, 54 (03) :559-566
[6]
Trueness verification of actual creatinine assays in the European market demonstrates a disappointing variability that needs substantial improvement An international study in the framework of the EC4 creatinine standardization working group [J].
Delanghe, Joris R. ;
Cobbaert, Christa ;
Galteau, Marie-Madeleine ;
Harmoinen, Aimo ;
Jansen, Rob ;
Kruse, Rolf ;
Laitinen, Paeivi ;
Thienpont, Linda M. ;
Wuyts, Birgitte ;
Weykamp, Cas ;
Panteghini, Mauro .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (09) :1319-1325
[7]
DELANGHE JR, 2008, NEPHROL DIAL TRANSPL
[8]
Extent of bilirubin interference with Beckman creatinine methods [J].
Dimeski, Goce ;
McWhinney, Brett ;
Jones, Brock ;
Mason, Rebecca ;
Carter, Andrew .
ANNALS OF CLINICAL BIOCHEMISTRY, 2008, 45 :91-92
[9]
*INT ORG STAND INT, 1993, GUID EXPR UNC MEAS
[10]
Khatami Z, 2005, ANN CLIN BIOCHEM, V42, P162