Assessment of multi-layer piezoelectric actuator technology for middle-ear implants

被引:48
作者
Wang, ZG
Abel, EW [1 ]
Mills, RP
Liu, Y
机构
[1] Univ Dundee, Med Engn Res Inst, Dundee DD1 4HN, Scotland
[2] Royal Infirm, Dept Otolaryngol, Edinburgh EH3 9EN, Midlothian, Scotland
[3] Univ Portsmouth, Dept Elect & Elect Engn, Portsmouth PO1 3DJ, Hants, England
关键词
piezoelectric actuator; hearing implant; temporal bone; vibration measurement;
D O I
10.1016/S0957-4158(00)00067-2
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A small mutti-layer piezoelectric actuator, which contains a stack of piezoelectric crystals in a thin-layered Compact Structure, was evaluated for its suitability as a stimulator in middle ear hearing implants. Using laser Doppler vibrometry for non-contact vibration measurement, the actuator was found to have a useful bandwidth spanning the range of audio frequencies. During simulated surgical implantation on human temporal bones, one side of the actuator was cemented to the medial wall of the attic of the middle ear and the other side was cemented to the body of the incus. Stapes displacement from actuator excitation at 3.5V(rms) was equivalent to that from acoustic stimulation at 90 dB SPL below 1 kHz, increasing to 120 dB at 8 kHz, which is more than adequate stimulation of the ossicular chain. The total harmonic distortion (THD) was less than 2.5%, over the audio frequency range. The corresponding power consumption was 0.7 mW per volt of excitation at I kHz, which is low enough for the actuator to be used in a partially implantable hearing implant. The surgical implantation and positioning of this type of actuator are straightforward due to the simple linear motion it provides. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:3 / 17
页数:15
相关论文
共 31 条
[1]   Audio-frequency characteristics of multilayer piezoelectric crystal actuator for use in hearing implant [J].
Abel, EW ;
Wang, ZG ;
Mills, RP ;
Liu, Y .
ELECTRONICS LETTERS, 2000, 36 (06) :494-495
[2]  
Ball G R, 1997, Ear Nose Throat J, V76, P213
[3]  
Ball G. R., 1997, ENT-EAR NOSE THROAT, V76, P222
[4]  
Ball G. R., 1997, ENT-EAR NOSE THROAT, V76, P220
[5]  
FREDERICKSON JM, 1995, OTOLARYNG CLIN N AM, V28, P107
[6]  
Garverick S L, 1997, Ear Nose Throat J, V76, P397
[7]  
GOODE RL, 1995, OTOLARYNG CLIN N AM, V28, P141
[8]  
Granstrom G, 1997, Ear Nose Throat J, V76, P238
[9]  
Gyo K., 1988, ADV AUDIOL, V4, P107
[10]  
Helfrick A. D., 1990, MODERN ELECT INSTRUM