Treatment head design for multileaf collimated high-energy electrons

被引:47
作者
Karlsson, MG [1 ]
Karlsson, M [1 ]
Ma, CM [1 ]
机构
[1] Umea Univ, Dept Radiat Phys, S-90185 Umea, Sweden
关键词
radiotherapy; electron beams; multileaf collimator; EGS4 Monte Carlo;
D O I
10.1118/1.598732
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper describes how a conventional treatment head can be modified for use of multileaf collimated electron beams. Automatic and dynamic beam delivery are possible for both electrons and photons by using the computer controlled multileaf collimator (MLC) for both photon and electron beams. Thereby, the electron beams can be mixed more freely into the treatment to take advantage of the specific depth modulation characteristics of electrons. The investigation was based on Monte Carlo calculations using the software package BEAM. The physical parameters used in this optimization were the beam penumbra and the virtual/effective point source position. These parameters are essential for shaping beams, beam matching and for dosimetry calculations. The optimization was carried out by modifying a number of parameters: replacing the air atmosphere in the treatment head with helium, adding a helium bag below the MLC, changing the position of the scattering foils, modifying the monitor chamber, and adjusting the position of the MLC. The beam characteristics for some of these designs were found to fulfil our criteria for clinically useful beams down to at least 9 MeV. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00610-0].
引用
收藏
页码:2161 / 2167
页数:7
相关论文
共 22 条
[1]  
ABOUMANDOUR M, 1978, STRAHLENTHER ONKOL, V154, P328
[2]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[3]   DESIGN PRINCIPLES AND CLINICAL POSSIBILITIES WITH A NEW GENERATION OF RADIATION-THERAPY EQUIPMENT - A REVIEW [J].
BRAHME, A .
ACTA ONCOLOGICA, 1987, 26 (06) :403-412
[4]  
BRAHME A, 7217 TRITAEPP ROYAL
[5]  
*IAEA, 1987, 277 IAEA
[6]   Radiotherapy of breast cancer after breast-conserving surgery:: an improved technique using mixed electron-photon beams with a multileaf collimator [J].
Jansson, T ;
Lindman, H ;
Nygård, K ;
Dahlgren, CV ;
Montelius, A ;
Öberg-Kreuger, C ;
Asplund, S ;
Bergh, J .
RADIOTHERAPY AND ONCOLOGY, 1998, 46 (01) :83-89
[7]   ELECTRON-BEAM CHARACTERISTICS OF THE 50-MEV RACETRACK MICROTRON [J].
KARLSSON, M ;
NYSTROM, H ;
SVENSSON, H .
MEDICAL PHYSICS, 1992, 19 (02) :307-315
[8]   MATCHING OF ELECTRON AND PHOTON BEAMS WITH A MULTI-LEAF COLLIMATOR [J].
KARLSSON, M ;
ZACKRISSON, B .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (03) :317-326
[9]   Exploration of new treatment modalities offered by high energy (up to 50 MeV) electrons and photons [J].
Karlsson, M ;
Zackrisson, B .
RADIOTHERAPY AND ONCOLOGY, 1997, 43 (03) :303-309
[10]   CLINICAL ELECTRON-BEAM DOSIMETRY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-25 [J].
KHAN, FM ;
DOPPKE, KP ;
HOGSTROM, KR ;
KUTCHER, GJ ;
NATH, R ;
PRASAD, SC ;
PURDY, JA ;
ROZENFELD, M ;
WERNER, BL .
MEDICAL PHYSICS, 1991, 18 (01) :73-109