β-particle-emitting radioactive stent implantation -: A safety and feasibility study

被引:34
作者
Wardeh, AJ
Kay, IP
Sabaté, M
Coen, VLMA
Gijzel, AL
Ligthart, JMR
den Boer, A
Levendag, PC
van der Giessen, WJ
Serruys, PW
机构
[1] Univ Hosp Dijkzigt, Ctr Heart, Thoraxctr, NL-3015 GD Rotterdam, Netherlands
[2] Dr Daniel Den Hoed Canc Ctr, NL-3008 AE Rotterdam, Netherlands
关键词
beta-rays; angioplasty; radioactive isotopes; restenosis; stents;
D O I
10.1161/01.CIR.100.16.1684
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-This study represents the Heart Center Rotterdam's contribution to the Isostents fur Restenosis Intervention Study, a nonrandomized multicenter trial evaluating the safety and feasibility of the radioactive Isostent in patients with single coronary artery disease. Restenosis after stent implantation is primarily caused by neointimal hyperplasia. In animal studies, beta-particle-emitting radioactive stents decrease neointimal hyperplasia by inhibiting smooth muscle cell proliferation. Methods non Results-The radioisotope P-32, a beta-particle emitter with a half-life of 14.3 days, was directly embedded into the Isostent. The calculated range of radioactivity was 0.75 to 1.5 mu Ci. Quantitative coronary angiography measurements were performed before and after the procedure and at 6-month follow-up. A total of 31 radioactive stents were used in 26 patients; 30 (97%) were successfully implanted, and 1 was embolized, Treated lesions were in the left anterior descending coronary artery (n=12), the right coronary artery (n=8), or the left circumflex coronary artery (n=6). Five patients received additional, nonradioactive stents. Treated lesion lengths were 13+/-4 mm, with a reference diameter of 2.93 +/- 0.47 mm. Minimum lumen diameter increased from 0.87+/-0.28 mm preprocedure to 2.84+/-0.35 mm postprocedure. No in-hospital adverse cardiac events occurred, All patients received aspirin indefinitely and ticlopidine for 4 weeks. Twenty-three patients (88%) returned for 6-month angiographic follow-up; 17% of them had in-stent restenosis, and 13% had repeat revascularization. No restenosis was observed at the stent edges, Minimum lumen diameter at follow-up averaged 1.85 +/- 0.69 mm, which resulted in a late loss of 0.99 +/- 0.59 mm and a late loss index of 0.53+/-0.35. No other major cardiac events occurred during the 6-month follow-up. Conclusions-The use of radioactive stents with an activity of 0.75 to 1.5 mu Ci is safe and feasible.
引用
收藏
页码:1684 / 1689
页数:6
相关论文
共 32 条
  • [1] Albiero R., 1998, European Heart Journal, V19, P457
  • [2] [Anonymous], 1994, QUANTITATIVE CORONAR
  • [3] THE ELECTROCARDIOGRAM IN POPULATION STUDIES - A CLASSIFICATION SYSTEM
    BLACKBURN, H
    KEYS, A
    SIMONSON, E
    RAUTAHARJU, P
    PUNSAR, S
    [J]. CIRCULATION, 1960, 21 (06) : 1160 - 1175
  • [4] Experimental results with endovascular irradiation via a radioactive stent
    Carter, AJ
    Laird, JR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (04): : 797 - 803
  • [5] Effects of endovascular radiation from a beta-particle-emitting stent in a porcine coronary restenosis model - A dose-response study
    Carter, AJ
    Laird, JR
    Bailey, LR
    Hoopes, TG
    Farb, A
    Fischell, DR
    Fischell, RE
    Fischel, TA
    Virmani, R
    [J]. CIRCULATION, 1996, 94 (10) : 2364 - 2368
  • [6] RESTENOSIS AFTER PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY - HAVE WE BEEN AIMING AT THE WRONG TARGET
    CURRIER, JW
    FAXON, DP
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (02) : 516 - 520
  • [7] CALIBRATION USING ANGIOGRAPHIC CATHETERS AS SCALING DEVICES - IMPORTANCE OF FILMING THE CATHETERS NOT FILLED WITH CONTRAST-MEDIUM
    DIMARIO, C
    HERMANS, WRM
    RENSING, BJ
    SERRUYS, PW
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (16) : 1377 - 1378
  • [8] SMALL STENT SIZE AND INTIMAL HYPERPLASIA CONTRIBUTE TO RESTENOSIS - A VOLUMETRIC INTRAVASCULAR ULTRASOUND ANALYSIS
    DUSSAILLANT, GR
    MINTZ, GS
    PICHARD, AD
    KENT, KM
    SATLER, LF
    POPMA, JJ
    WONG, SC
    LEON, MB
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) : 720 - 724
  • [9] A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE
    FISCHMAN, DL
    LEON, MB
    BAIM, DS
    SCHATZ, RA
    SAVAGE, MP
    PENN, I
    DETRE, K
    VELTRI, L
    RICCI, D
    NOBUYOSHI, M
    CLEMAN, M
    HEUSER, R
    ALMOND, D
    TEIRSTEIN, PS
    FISH, RD
    COLOMBO, A
    BRINKER, J
    MOSES, J
    SHAKNOVICH, A
    HIRSHFELD, J
    BAILEY, S
    ELLIS, S
    RAKE, R
    GOLDBERG, S
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) : 496 - 501
  • [10] Extracellular matrix collagen synthesis and degradation following coronary balloon angioplasty
    Guarda, E
    Katwa, LC
    Campbell, SE
    Tanner, MA
    Webel, RM
    Laughlin, H
    Jenkins, S
    Myers, PR
    [J]. JOURNAL OF MOLECULAR AND CELLULAR CARDIOLOGY, 1996, 28 (04) : 699 - 706