Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry

被引:177
作者
Fetics, B
Nevo, E
Chen, CH
Kass, DA [1 ]
机构
[1] Johns Hopkins Med Inst, Dept Med, Div Cardiol, Baltimore, MD 21287 USA
[2] Johns Hopkins Med Inst, Dept Biomed Engn, Baltimore, MD 21287 USA
[3] Natl Yang Ming Med Coll, Taipei 11221, Taiwan
[4] Vet Gen Hosp, Taipei, Taiwan
关键词
blood pressure; tonometry; transfer function;
D O I
10.1109/10.764946
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aortic pressure can be estimated noninvasively by applying a transfer function (TF) to radial tonometry signals. This study compares the performance of prior approaches, based on Fourier transform and inverted aortic-to-radial model, with direct radial-to-aortic autoregressive exogenous (ARX) model, Simultaneous invasive aortic pressure and radial tonometry pressure were recorded during rest in 39 patients in the supine position, Individual radial-aortic TF's were estimated from 20 patients, and the average TF was used to predict aortic pressures in the remaining 19 patients. The direct average TF yielded accurate aortic systolic pressure estimation (error 0.4+/-2.9 mmHg) and good reproduction of the aortic pressure waveform (root mean squared error 2.2+/-0.9 mmHg), The inverted reverse TF (aortic radial) yielded comparable results, while the Fourier-based TF had worse performance. Individual direct TF provided improved predictive accuracy only for indexes which are based on higher frequency components of the waveform (augmentation index, systolic time period). An ARX average TF can be used to accurately estimate central aortic pressure waveform parameters from noninvasive radial pulse tracings, and its performance is superior to previous techniques.
引用
收藏
页码:698 / 706
页数:9
相关论文
共 23 条
  • [1] Reconstruction of brachial artery pressure from noninvasive finger pressure measurements
    Bos, WJW
    vanGoudoever, J
    vanMontfrans, GA
    vandenMeiracker, AH
    Wesseling, KH
    [J]. CIRCULATION, 1996, 94 (08) : 1870 - 1875
  • [2] AUTOREGRESSIVE SPECTRAL MODELS OF HEART-RATE-VARIABILITY - PRACTICAL ISSUES
    BURR, RL
    COWAN, MJ
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 1993, 25 : 224 - 233
  • [3] Validation of carotid artery tonometry as a means of estimating augmentation index of ascending aortic pressure
    Chen, CH
    Ting, CT
    Nussbacher, A
    Nevo, E
    Kass, DA
    Pak, P
    Wang, SP
    Chang, MS
    Yin, FCP
    [J]. HYPERTENSION, 1996, 27 (02) : 168 - 175
  • [4] Chen CH, 1997, CIRCULATION, V95, P1827
  • [5] PRESSURE-DEPENDENCE OF MECHANICAL-PROPERTIES OF ARTERIES INVIVO
    COX, RH
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1975, 229 (05): : 1371 - 1375
  • [6] AORTOFEMORAL TRANSFER-FUNCTION - A METHOD TO DETERMINE THE INSTANTANEOUS AORTIC-VALVE GRADIENT IN AORTIC-VALVE STENOSIS
    FITCHETT, DH
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (07) : 1909 - 1914
  • [7] AN ANALYSIS OF THE RELATIONSHIP BETWEEN CENTRAL AORTIC AND PERIPHERAL UPPER LIMB PRESSURE WAVES IN MAN
    KARAMANOGLU, M
    OROURKE, MF
    AVOLIO, AP
    KELLY, RP
    [J]. EUROPEAN HEART JOURNAL, 1993, 14 (02) : 160 - 167
  • [8] PRESSURE WAVE-PROPAGATION IN A MULTIBRANCHED MODEL OF THE HUMAN UPPER-LIMB
    KARAMANOGLU, M
    GALLAGHER, DE
    AVOLIO, AP
    OROURKE, MF
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 269 (04): : H1363 - H1369
  • [9] PRESSURE-VOLUME ANALYSIS AS A METHOD FOR QUANTIFYING SIMULTANEOUS DRUG (AMRINONE) EFFECTS ON ARTERIAL LOAD AND CONTRACTILE STATE INVIVO
    KASS, DA
    GRAYSON, R
    MARINO, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) : 726 - 732
  • [10] SPECTRUM ANALYSIS - A MODERN PERSPECTIVE
    KAY, SM
    MARPLE, SL
    [J]. PROCEEDINGS OF THE IEEE, 1981, 69 (11) : 1380 - 1419