Femoral component wear in retrieved hip prostheses

被引:35
作者
Kusaba, A [1 ]
Kuroki, Y [1 ]
机构
[1] SHOWA UNIV,FUJIGAOKA REHABIL HOSP,DEPT ORTHOPAED SURG,AOBA KU,YOKOHAMA,KANAGAWA 227,JAPAN
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1997年 / 79B卷 / 02期
关键词
D O I
10.1302/0301-620X.79B2.7217
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We retrieved 159 femoral heads at revision surgery to determine changes in surface configuration. Macroscopic wear of the head was observed in three bipolar hip prostheses as a result of three-body wear. There was a considerable change in surface roughness in the internal articulation of bipolar hip prostheses. Roughness in alumina heads was almost the same as that in new cobalt-chromium heads. The annual linear wear rate of polyethylene cups with alumina heads was less than that of cups with cobalt-chromium alloy heads. Polyethylene wear was increased in the prostheses which had increased roughness of the head.
引用
收藏
页码:331 / 336
页数:6
相关论文
共 24 条
[1]  
BLACK J, 1988, ORTHOPAEDIC BIOMATER, P319
[2]  
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[3]  
Charnley J, 1979, LOW FRICTION ARTHROP, P320
[4]   CHARACTERISTICS OF ACETABULAR CUPS WORN IN HUMAN-BODY [J].
DOWLING, JM ;
ATKINSON, JR ;
DOWSON, D ;
CHARNLEY, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :375-382
[5]  
DUMBLETON JH, 1984, COMPLICATIONS TOTAL, P212
[6]  
Feller J A, 1994, J Arthroplasty, V9, P341, DOI 10.1016/0883-5403(94)90042-6
[7]  
GRIFFITH MJ, 1978, CLIN ORTHOP RELAT R, V137, P37
[8]  
HERNANDEZ JR, 1994, J BONE JOINT SURG BR, V76B, P263
[9]   INVIVO WEAR OF POLYETHYLENE ACETABULAR COMPONENTS [J].
KABO, JM ;
GEBHARD, JS ;
LOREN, G ;
AMSTUTZ, HC .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02) :254-258
[10]   LOW WEAR RATE OF UHMWPE AGAINST ZIRCONIA CERAMIC (Y-PSZ) IN COMPARISON TO ALUMINA CERAMIC AND SUS 316L ALLOY [J].
KUMAR, P ;
OKA, M ;
IKEUCHI, K ;
SHIMIZU, K ;
YAMAMURO, T ;
OKUMURA, H ;
KOTOURA, Y .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (07) :813-828