Electrohydrodynamic Jetting Behaviour of Polyhedral Oligomeric Silsesquioxane Nanocomposite

被引:18
作者
Bakhshi, R. [1 ,2 ]
Edirisinghe, M. J. [1 ]
Darbyshire, A. [2 ]
Ahmad, Z. [1 ]
Seifalian, A. M. [2 ]
机构
[1] UCL, Dept Mech Engn, London WC1E 7JE, England
[2] UCL, Acad Div Surg & Intervent Sci, Biomat & Tissue Engn Ctr, London NW3 2PF, England
基金
英国工程与自然科学研究理事会;
关键词
electrohydrodynamic spraying; stent; coating; nanocomposite; polymer; silsesquioxane; DRUG-ELUTING STENTS; ELECTROSTATIC SPRAY DEPOSITION; STEEL CORONARY STENTS; IN-VITRO; CARDIOVASCULAR STENTS; BIOMEDICAL APPLICATIONS; DIELECTRIC LIQUIDS; METALLIC STENTS; COATINGS; RELEASE;
D O I
10.1177/0885328208089125
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, we investigate in detail the electrohydrodynamic spraying of a nonbiodegradable nanocomposite polyhedral oligomeric silsesquioxane polymer developed in our laboratories and currently being explored for coating metallic stent materials. Different concentrations of the polymer have been dissolved to prepare, characterise, and electrohydrodynamically deposit the polymer on stainless steel. From the experiments, the solution containing 15 wt% polymer was selected for further investigation. The variation of film/coating thickness as a function of spraying time was studied and the structural features of the film were assessed using microscopy. Films were also tensile tested. This study has identified a process and conditions which can be used in our stent coating research.
引用
收藏
页码:293 / 309
页数:17
相关论文
共 55 条
[1]  
Ajani Andrew E, 2007, Heart Lung Circ, V16, P269, DOI 10.1016/j.hlc.2007.02.089
[2]   The future of drug eluting stents [J].
Anis, RR ;
Karsch, KR .
HEART, 2006, 92 (05) :585-588
[3]   Biocompatibility aspects of new stent technology [J].
Bertrand, OF ;
Sipehia, R ;
Mongrain, R ;
Rodés, JR ;
Tardif, JC ;
Bilodeau, L ;
Côté, G ;
Bourassa, MG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (03) :562-571
[4]   A novel drug-eluting stent coated with an integrin-binding cyclic Arg-Gly-Asp peptide inhibits neointimal hyperplasia by recruiting endothelial progenitor cells [J].
Blindt, R ;
Vogt, F ;
Astafieva, I ;
Fach, C ;
Hristov, M ;
Krott, N ;
Seitz, B ;
Kapurniotu, A ;
Kwok, C ;
Dewor, M ;
BosserhofF, AK ;
Bernhagen, J ;
Hanrath, P ;
Hofmann, R ;
Weber, C .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (09) :1786-1795
[5]   Late coronary thrombosis in paclitaxel-eluting stents. Case reports [J].
Cassin, Matteo ;
De Biasio, Marzia ;
Macor, Franco ;
Burelli, Claudio ;
Vendrametto, Fauzia ;
Driussi, Mauro ;
Nicolosi, Gian Luigi .
JOURNAL OF CARDIOVASCULAR MEDICINE, 2007, 8 (04) :284-288
[6]   Morphology control of thin LiCoO2 films fabricated using the electrostatic spray deposition (ESD) technique [J].
Chen, CH ;
Kelder, EM ;
vanderPut, PJJM ;
Schoonman, J .
JOURNAL OF MATERIALS CHEMISTRY, 1996, 6 (05) :765-771
[7]   Plasticity as a lifesaver in the design of cardiovascular stents [J].
De Beule, Matthieu ;
Mortier, Peter ;
Belis, Jan ;
Van Impe, Rudy ;
Verhegghe, Benedict ;
Verdonck, Pascal .
ENGINEERING PLASTICITY AND ITS APPLICATIONS FROM NANOSCALE TO MACROSCALE, PTS 1 AND 2, 2007, 340-341 :841-+
[8]   BIOCOMPATIBILITY OF POLYMER-COATED OVERSIZED METALLIC STENTS IMPLANTED IN NORMAL PORCINE CORONARY-ARTERIES [J].
DESCHEERDER, IK ;
WILCZEK, KL ;
VERBEKEN, EV ;
VANDORPE, J ;
LAN, PN ;
SCHACHT, E ;
DEGEEST, H ;
PIESSENS, J .
ATHEROSCLEROSIS, 1995, 114 (01) :105-114
[9]   Bioresorbable polymeric stents: current status and future promise [J].
Eberhart, RC ;
Su, SH ;
Nguyen, KT ;
Zilberman, M ;
Tang, LP ;
Nelson, KD ;
Frenkel, P .
JOURNAL OF BIOMATERIALS SCIENCE-POLYMER EDITION, 2003, 14 (04) :299-312
[10]  
Edirisinghe MJ, 2004, INT J APPL CERAM TEC, V1, P140