Efficacy and complications of Angioseal in antegrade puncture

被引:26
作者
Mukhopadhyay, K [1 ]
Puckett, MA [1 ]
Roobottom, CA [1 ]
机构
[1] Derriford Hosp, Plymouth PL6 8DH, Devon, England
关键词
interventional procedures; angiography; haemorrhage; cost-effectiveness;
D O I
10.1016/j.ejrad.2005.03.025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: Although arterial closure devices are in use for over a decade, there have been no specific study to assess the use of the same in antegrade puncture. Material and method: Consecutive patients undergoing antegrade puncture for peripheral vascular disease in a single centre performed by a single operator were included in this study. The notes and the angiograms were reviewed retrospectively for possible complications and the severity of disease. Result: Over a 25-month period, 21 patients with antegrade puncture had Angioseal deployed for haemostasis. Post-procedure complications recorded from the case notes showed one episode of haematoma formation and one case of worsening ischaemia. The severity of the disease at the puncture site did not influence the outcome. Conclusion: Use of Angioseal is safe and effective in antegrade punctures for peripheral vascular disease. The result is independent of the severity of disease at the puncture site. (C) 2005 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:409 / 412
页数:4
相关论文
共 14 条
[1]  
Carey D, 2001, CATHETER CARDIO INTE, V52, P3, DOI 10.1002/1522-726X(200101)52:1<3::AID-CCD1002>3.0.CO
[2]  
2-G
[3]   Effect of a closure device on complication rates in high-local-risk patients: Results of a randomized multicenter trial [J].
Chevalier, B ;
Lancelin, B ;
Koning, R ;
Henry, M ;
Gommeaux, A ;
Pilliere, R ;
Elbaz, M ;
Lefevre, T ;
Boughalem, K ;
Marco, J ;
Dupouy, P .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 58 (03) :285-291
[4]   Suture-mediated percutaneous closure of antegrade femoral arterial access sites in patients who have received full anticoagulation therapy [J].
Duda, SH ;
Wiskirchen, J ;
Erb, M ;
Schott, U ;
Khaligi, K ;
Pereira, PL ;
Albes, J ;
Claussen, CD .
RADIOLOGY, 1999, 210 (01) :47-52
[5]  
Duffin D C, 2001, J Invasive Cardiol, V13, P354
[6]   Complications associated with percutaneous closure devices [J].
Gonze, MD ;
Sternbergh, WC ;
Salartash, K ;
Money, SR .
AMERICAN JOURNAL OF SURGERY, 1999, 178 (03) :209-211
[7]  
KENSEY KR, 1994, J INVASIVE CARDIOL, V6, P273
[8]   Local vascular complications after use of the hemostatic puncture closure device Angio-Seal™ [J].
Kirchhof, C ;
Schickel, S ;
Schmidt-Lucke, C ;
Schmidt-Lucke, JA .
VASA-JOURNAL OF VASCULAR DISEASES, 2002, 31 (02) :101-106
[9]   A prospective randomized trial comparing the safety and efficacy of three commercially available closure devices (Angioseal, Vasoseal and Duett) [J].
Michalis, LK ;
Rees, MR ;
Patsouras, D ;
Katsouras, CS ;
Goudevenos, J ;
Pappas, S ;
Sourla, E ;
Kolettis, T ;
Sioros, L ;
Zotou, P ;
Gartzou-Matsouka, P ;
Sideris, DA .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 25 (05) :423-429
[10]   The Angio-Seal™ hemostatic puncture closure device -: Concept and experimental results [J].
Nash, JE ;
Evans, DG .
HERZ, 1999, 24 (08) :597-606