Comparison of wear in a total knee replacement under different kinematic conditions

被引:49
作者
Barnett, PI
Fisher, J
Auger, DD
Stone, MH
Ingham, E
机构
[1] Univ Leeds, Dept Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] DePuy Int, Leeds LS11 8DT, W Yorkshire, England
[3] Gen Infirm, Dept Orthopaed Surg, Leeds LS1 3EX, W Yorkshire, England
[4] Univ Leeds, Old Med Sch, Div Microbiol, Leeds LS2 9NL, W Yorkshire, England
关键词
Polyethylene; Wear Rate; Axial Load; Abrasive Wear; Knee Replacement;
D O I
10.1023/A:1012894023657
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
A six station ProSim (Manchester, UK) knee simulator was used to assess the wear of six PFC (DePuy) fixed bearing total knee replacements under two different kinematic conditions defined as low and high kinematic inputs. The high kinematics displacement and rotation inputs were based on the kinematics of the natural knee with ISO standards used for the axial load and flexion. Low kinematics were defined as approximately half the magnitude. The six specimens were run for three million cycles under low kinematics and three million cycles under high kinematics. The mean wear rate found during the low kinematics phase was 7.7 +/- 2mm(3) per million cycles. This then increased significantly to an average wear rate of 41 +/- 14mm(3) during the high kinematics input phase. The wear areas were characterized by a predominant damage mode of burnishing with some abrasive wear occurring during the high kinematics phase. This study supports the findings that introduction of cross-shearing of the polyethylene by introducing both rotational and anterior/posterior displacement increases the wear rate. This has implications for younger patients with higher levels of activity that need knee replacements. (C) 2001 Kluwer Academic Publishers.
引用
收藏
页码:1039 / 1042
页数:4
相关论文
共 8 条
[1]
Development of a six station knee wear simulator and preliminary wear results [J].
Burgess, IC ;
Kolar, M ;
Cunningham, JL ;
Unsworth, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :37-47
[2]
PRELIMINARY-STUDY OF THE EFFECT OF AGING FOLLOWING IRRADIATION ON THE WEAR OF ULTRAHIGH-MOLECULAR-WEIGHT POLYETHYLENE [J].
FISHER, J ;
CHAN, KL ;
HAILEY, JL ;
SHAW, D ;
STONE, M .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :689-692
[3]
FURMAN BD, 1999, P 45 ANN M ORTH RES, P149
[4]
GILLIS AM, 2000, T 6 WORLD BIOM C HAW, P555
[5]
JOHNSON TS, 2000, T 6 WORLD BIOM C HAW, P56
[6]
3-DIMENSIONAL KINEMATICS OF THE HUMAN KNEE DURING WALKING [J].
LAFORTUNE, MA ;
CAVANAGH, PR ;
SOMMER, HJ ;
KALENAK, A .
JOURNAL OF BIOMECHANICS, 1992, 25 (04) :347-357
[7]
SCHMIDIG G, 2000, 46 ORS ORL FL, P555
[8]
Wear testing of UHMWPE tibial components: Influence of oxidation [J].
Young, SK ;
Keller, TS ;
Greer, KW ;
Gorhan, MC .
JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2000, 122 (01) :323-331