Has conventional sphygmomanometry ended with the banning of mercury?

被引:30
作者
O'Brien, E [1 ]
机构
[1] Beaumont Hosp, Blood Pressure Unit, Dublin 9, Ireland
[2] Beaumont Hosp, ADAPT Ctr, Dublin 9, Ireland
关键词
blood pressure measurement; mercury sphygmomanometer; automated devices; European Society of Hypertension; device validation; kilopascal; millimetre of mercury;
D O I
10.1097/00126097-200202000-00007
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The banning of mercury from clinical practice will lead to the inevitable demise of traditional clinical sphygmomanometry. There are differences in approach to this important issue between European countries on the one hand, which generally have accepted that the mercury sphygmomanometer must be replaced with alternative devices, and the U.S. on the other, where the view is that the mercury sphygmomanometer should remain as the mainstay of blood pressure measurement. The availability of alternative devices for the mercury sphygmomanometer is improving but the problem of independent validation is a serious issue, which is being addressed by the European Society of Hypertension Working Party on Blood Pressure Monitoring, which has drafted an International Protocol for validating blood pressure measuring devices. The removal of the mercury sphygmomanometer from clinical practice has other implications, which merit careful consideration; the advent of automated devices must lead inevitably to the disappearance of the traditional clinical auscultatory technique of blood pressure measurement, and with the disappearance of mercury it will be argued that the Systeme Intemational(SI) unit of measurement - the kilopascal - should replace the millimetre of mercury. Blood Pressure Monit 7:37-40 (C) 2002 Lippincott Williams Wilkins.
引用
收藏
页码:37 / 40
页数:4
相关论文
共 22 条
[1]  
*AM NAT STAND, 1993, EL AUT SPHYGM, P22201
[2]  
*EUR COMM STAND, 1994, 62 CENELECTC EUR COM
[3]   Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor [J].
Jones, CR ;
Taylor, K ;
Poston, L ;
Shennan, AH .
JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (03) :191-195
[4]   Mercury sphygmomanometers should not be abandoned: An advisory statement from the council for high blood pressure research, American Heart Association [J].
Jones, DW ;
Frohlich, ED ;
Grim, CM ;
Grim, CE ;
Taubert, KA .
HYPERTENSION, 2001, 37 (02) :185-186
[5]  
Jones DW, 2001, HYPERTENSION, V38, pE20
[6]   What unit for blood pressure? [J].
MacGregor, G .
LANCET, 2001, 357 (9266) :1452-1452
[7]   The mercury sphygmomanometer should be abandoned before it is proscribed [J].
Markandu, ND ;
Whitcher, F ;
Arnold, A ;
Carney, C .
JOURNAL OF HUMAN HYPERTENSION, 2000, 14 (01) :31-36
[8]  
MARKANDU ND, 2001, J HUM HYPERTENS, V15, P357
[9]  
*MED ADV AG, 2000, MED ADV AG MERC DEV, V3, P1
[10]  
O'Brien, 1996, Blood Press Monit, V1, P55