Cochlear implant deep electrode insertion: Extent of insertional trauma

被引:79
作者
Gstoettner, W
Plenk, H
Franz, P
Hamzavi, J
Baumgartner, W
Czerny, C
Ehrenberger, K
机构
[1] UNIV VIENNA,DEPT OTOLARYNGOL,VIENNA,AUSTRIA
[2] UNIV VIENNA,INST HISTOL EMBRYOL,BONE & BIOMAT RES DEPT,A-1010 VIENNA,AUSTRIA
关键词
hearing nerve; auditory; electrostimulation; human temporal bones; Combi; 40;
D O I
10.3109/00016489709117786
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
We have recently undertaken deep insertions of the Combi-40 cochlear implant electrode (Med-Er Corp., Innsbruck, Austria) into apical regions of the scala tympani using a cochleostomy approach. In order to examine the extent of the insertional trauma, 12 fresh human temporal bones were implanted with original Combi-40 electrodes. The specimens were histologically processed with the implants in place by employing a sawing and grinding technique. In most cases, only very discrete distortions of the epithelium of the spiral ligament occurred within the middle cochlear turns. Furthermore, a slight displacement of the basilar membrane caused by the electrode was occasionally seen. However, in 2 cases more severe damage such as basilar membrane rupture and electrode displacement was found. Attempts to insert the electrode beyond the point of first resistance resulted in electrode kinking within the basal cochlear turn with subsequent fracture of the osseous spiral lamina. According to our results, deep electrode insertions do not aggravate the insertional trauma provided no force is applied when resistance is felt.
引用
收藏
页码:274 / 277
页数:4
相关论文
共 14 条
[1]  
Donnely MJ, 1995, AUST J OTOLARYNGOL, V1, P79
[2]   COCHLEAR IMPLANTS - HISTOPATHOLOGIC FINDINGS RELATED TO PERFORMANCE IN 16 HUMAN TEMPORAL BONES [J].
FAYAD, J ;
LINTHICUM, FH ;
OTTO, SR ;
GALEY, FR ;
HOUSE, WF .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1991, 100 (10) :807-811
[3]   ANATOMICAL PERSPECTIVE, APPROACH, AND EXPERIENCE WITH MULTICHANNEL INTRACOCHLEAR IMPLANTATION [J].
GOYCOOLEA, MV ;
MUCHOW, DC ;
SCHIRBER, CM ;
GOYCOOLEA, HG ;
SCHELLHAS, K .
LARYNGOSCOPE, 1990, 100 (02) :1-18
[4]  
GSTOETTNER W, 1996, IN PRESS LARYNGOSCOP
[5]  
GSTOETTNER W, 1996, CENTRAL E EUROPEAN J, V1, P214
[6]  
KENNEDY DW, 1987, LARYNGOSCOPE, V97, P42
[7]  
LENHARDT E, 1993, COCHLEAR IMPLANTS NE, V48, P62
[8]  
MARSH MA, 1992, AM J OTOL, V13, P241
[9]  
NADOL JB, 1994, LARYNGOSCOPE, V104, P299
[10]   ELECTRODE INSERTION TRAUMA IN COCHLEAR IMPLANTATION [J].
OLEARY, MJ ;
HOUSE, WF ;
FAYAD, J ;
LINTHICUM, FH .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1991, 100 (09) :695-699