A new Zn(II)-selective potentiometric sensor based on 4-tert-butylcalix[4]arene in PVC matrix

被引:39
作者
Gupta, V. K. [1 ]
Goyal, R. N.
Al Khayat, M.
Kumar, P.
Bachheti, N.
机构
[1] Indian Inst Technol, Dept Chem, Roorkee 247667, Uttar Pradesh, India
[2] Univ Sharjah, Dept Basic Sci, Sharjah, U Arab Emirates
[3] Cent Bldg Res Inst, Environm Sci & Technol Div, Roorkee 247667, Uttar Pradesh, India
关键词
ion-selective electrodes; sensors; zinc sensors; waste water;
D O I
10.1016/j.talanta.2005.12.040
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Poly(vinyl chloride) (PVC) based membranes containing 4-tert-butylcalix[4]arene (I) as an electroactive material alongwith anion excluder sodiumtetraphenylborate (NaTPB) and plasticizer tri-butylphosphate (TBP) have been developed to fabricate a new zinc-selective sensor. Out of various compositions, the best performance was exhibited by the membrane having I, NaTPB, TBP and PVC in the ratio 8:5:100:200 (w/w). The sensor works well in the concentration range 9.8 x 10(-6) to 1.0 x 10(-1) mol dm(-3) with a near-Nernstian slope of 28.0 +/- 1.0 mV/decade of activity. The detection limit is down to 5.0 x 10(-7) mol dm(-3). The working pH range of this sensor is 2.5-4.3 and it works well in partially non-aqueous medium up to 15% (v/v) (methanol, ethanol and acetone). It exhibits a fast response time of 30 s and could be used for more than four months without any considerable change in response characteristics. It has excellent selectivity for Zn(II) over other mono-, bi- and trivalent cations which have been reported to cause interference in the working of other sensors. It has been successfully used as an indicator electrode in the potentiometric titration of Zn(II) against EDTA and also to estimate zinc ions in industrial waste waters. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:1149 / 1155
页数:7
相关论文
共 35 条
[1]  
AMMAN D, 1991, ANAL CHIM ACTA, V171, P1380
[2]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[3]  
Bouen S., 1982, J CHEM SOC CHEM COMM, V20, P1172
[4]   DETERMINATION AND APPLICATION OF ION-SELECTIVE ELECTRODE MODEL PARAMETERS USING FLOW-INJECTION AND SIMPLEX OPTIMIZATION [J].
DEVITERI, FJS ;
DIAMOND, D .
ANALYST, 1994, 119 (05) :749-758
[5]   Zn(II)-selective membrane electrode based on tetra(2-aminophenyl) porphyrin [J].
Fakhari, AR ;
Shamsipur, M ;
Ghanbari, K .
ANALYTICA CHIMICA ACTA, 2002, 460 (02) :177-183
[6]   DETERMINATION OF SELECTIVITY COEFFICIENTS OF ION-SELECTIVE ELECTRODES BY A MATCHED-POTENTIAL METHOD [J].
GADZEKPO, VPY ;
CHRISTIAN, GD .
ANALYTICA CHIMICA ACTA, 1984, 164 (OCT) :279-282
[7]   PVC-based bis(2-nitrophenyl)disulfide sensor for zinc ions [J].
Gholivand, MB ;
Mozaffari, Y .
TALANTA, 2003, 59 (02) :399-407
[8]  
Gokel G.W., 1991, CROWN ETHERS CRYPTAN
[9]   ZINC-SENSITIVE POLYMERIC MEMBRANE ELECTRODE [J].
GORTON, L ;
FIEDLER, U .
ANALYTICA CHIMICA ACTA, 1977, 90 (MAY) :233-236
[10]   Porphyrins as carrier in PVC based membrane potentiometric sensors for nickel(II) [J].
Gupta, VK ;
Jain, AK ;
Singh, LP ;
Khurana, U .
ANALYTICA CHIMICA ACTA, 1997, 355 (01) :33-41