Analysis of feline, canine and equine hemograms using the QBC VetAutoread

被引:20
作者
Papasouliotis, K [1 ]
Cue, S [1 ]
Graham, M [1 ]
Sparkes, AH [1 ]
Gruffydd-Jones, T [1 ]
机构
[1] Univ Bristol, Dept Vet Clin Sci, Div Compan Anim, Bristol BS18 7DU, Avon, England
关键词
canine; equine; feline; hemograms; QBC VetAutoread;
D O I
10.1111/j.1939-165X.1999.tb01058.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Blood samples from 120 consecutive clinical cases (40 cats, 40 dogs and 40 horses) were analyzed on the QBC VetAutoread analyzer and the results compared with those obtained by a Baker 9000 electronic resistance cell counter and a 100-cell manual differential leukocyte (WBC) count. Packed cell volume (PCV) hemoglobin (Hb) concentration, mean cell hemoglobin concentration (MCHC), and platelet, total WBC, granulocyte, and lymphocyte plus monocyte (L+M) counts were determined. Indistinct separation of red blood cell and granulocyte layers on the QBC VetAutoread was observed in samples from five cats (12.5%), two dogs (5%), and one horse. Significantly different (P = 0.002) median values for the two methods were obtained for PCV, Hb concentration, MCHC and platelet count in cats; PCV, MCHC, WBC count and granulocyte count in dogs; and PCV, Hb concentration, MCHC, and WBC, granulocyte and platelet counts in horses. Results fi om the QBC VetAutoread should not be interpreted using reference ranges established using other equipment. Results were abnormal on a limited number of samples; however, when correlation coefficients were low, marked discrepancy existed between values within Its well as outside of reference ranges. Spearman rank correlation coefficients were excellent (r = 0.93) for PCV and Hb concentration in dogs, and Hb concentration and WBC count in horses. Correlation was good (r = 0.80-0.92) for PCV and Hb concentration in cats, WBC count in dogs, and PCV, granulocyte count and platelet count in horses. For remaining parameters, correlation was fair to poor (r = 0.79). Acceptable correlations (r > 0.80) were achieved between the two test systems for all equine values except MCHC and L+M count, but only for PCV and Hb concentration in feline and canine blood samples.
引用
收藏
页码:109 / 115
页数:7
相关论文
共 12 条
[1]  
BROWN SA, 1988, AM J VET RES, V49, P321
[2]  
Fischer A, 1989, Tierarztl Prax, V17, P227
[3]  
Guelfi J. F., 1989, Pratique Medicale & Chirurgicale de l'Animal de Compagnie, V24, P485
[4]  
Guelfi JF, 1995, REV MED VET-TOULOUSE, V146, P765
[5]   Clinical hematology - In-clinic analysis, quality control, reference values, and system selection [J].
Knoll, JS ;
Rowell, SL .
VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 1996, 26 (05) :981-&
[6]  
LEVINE RA, 1986, J AM VET MED ASSOC, V189, P670
[7]  
NAKEFF A, 1961, J EXP MED, V14, P51
[8]  
Tvedten H., 1994, Small animal clinical diagnosis by laboratory methods., P11
[9]  
Tvedten H W, 1988, Vet Clin Pathol, V17, P47, DOI 10.1111/j.1939-165X.1988.tb00490.x
[10]   QUANTITATIVE BUFFY COAT ANALYSIS - A NEW LABORATORY TOOL FUNCTIONING AS A SCREENING COMPLETE BLOOD-CELL COUNT [J].
WARDLAW, SC ;
LEVINE, RA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1983, 249 (05) :617-620