Determination of cyclosporin A in 20% ethanol by a magnetic beads-based immunofluorescence assay

被引:22
作者
Kiselev, MV
Gladilin, AK [1 ]
Melik-Nubarov, NS
Sveshnikov, PG
Miethe, P
Levashov, AV
机构
[1] Moscow MV Lomonosov State Univ, Dept Chem, Moscow 119899, Russia
[2] ABION, Gesselschaft Biowissensch & Tech mbH, D-52428 Juelich, Germany
关键词
D O I
10.1006/abio.1999.4015
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid magnetic beads-based immunoassay for the immunodepressant drug cyclosporin A (CsA) has been developed. The method allows CsA determination in medium with a higher content of ethanol compared to conventional immunochemical techniques due to increased antibody stability. Monitoring of the drug in ethanol extracts from patient's whole blood without many-fold dilution with aqueous buffer is possible. The assay has adequate specificity and sensitivity for CsA to be suitable for the routine monitoring of therapy, (C) 1999 Academic Press.
引用
收藏
页码:393 / 398
页数:6
相关论文
共 28 条
[1]  
*ABB LAB, 1988, INS TDX CYCL MET WHO
[2]   Effect of organic solvents on lysozyme-antilysozyme precipitin reaction [J].
Ahmad, A ;
Salahuddin, A .
COMPARATIVE BIOCHEMISTRY AND PHYSIOLOGY C-PHARMACOLOGY TOXICOLOGY & ENDOCRINOLOGY, 1996, 114 (02) :119-121
[3]   Water immiscible solvent based immunoassay [J].
Aston, JP ;
Hitchings, EJ ;
Ball, RL ;
Weeks, I ;
Woodhead, JS .
JOURNAL OF IMMUNOASSAY, 1997, 18 (03) :235-246
[4]  
BAUMANN G, 1992, IMMUNOL TODAY, V13, P136
[5]  
BOREL JF, 1989, PHARMACOL REV, V41, P239
[6]   ANTIBODIES TO CYCLOSPORINE-A (CSA) BY A NOVEL ROUTE AND THEIR USE TO MONITOR CYCLOSPORINE LEVELS BY RADIOIMMUNOASSAY (RIA) [J].
CACALANO, NA ;
CLEVELAND, WL ;
ERLANGER, BF .
JOURNAL OF IMMUNOLOGICAL METHODS, 1989, 118 (02) :257-263
[7]  
COLLINS W, COMPLEMENTARY IMMUNO
[8]  
Gladilin AK, 1995, BIOTECHNOL LETT, V17, P1329
[9]  
GREVEL J, 1988, TRANSPLANT P, V20, P428
[10]   Whole blood cyclosporin monitoring in liver and heart transplant patients: Evaluation of the specificity of a fluorescence polarization immunoassay and an enzyme-multiplied immunoassay technique [J].
Gulbis, B ;
VanderHeijden, J ;
vanAs, H ;
Thiry, P .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1997, 15 (07) :957-963