Prevention of water vapour adsorption by carbon molecular sieves in sampling humid gases

被引:36
作者
Gawrys, M [1 ]
Fastyn, P [1 ]
Gawlowski, J [1 ]
Gierczak, T [1 ]
Niedzielski, J [1 ]
机构
[1] Univ Warsaw, Dept Chem, PL-02089 Warsaw, Poland
关键词
air analysis; instrumentation; sample preparation; molecular sieves; Dubinin-Astakhov isotherm; adsorption isotherms; volatile organic compounds;
D O I
10.1016/S0021-9673(01)01250-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The water uptake by the solid sorbents Carbosieve S-III, Carboxen 569, 1000 and 1001, all of which are used for sampling of volatile organic compounds from the atmosphere, was examined using a direct experimental approach. The content of retained water is affected both by the trap temperature and the initial water vapour concentration in the sampled gas. Two different adsorption mechanism are operative. At low relative humidities (RH) only active polar centres are involved. This adsorption is so weak that negative water interferences can easily be managed. Another mechanism, the micropore volume filling, involves substantial amounts of water, becomes operative once the threshold value for relative humidity (RHth) is surpassed. RHth is 45 +/-3% for Carboxen 1000 decreasing to 35 +/-3% for the three other sorbents studied. A novel but simple strategy was tested for water management: moderate heating of the trap during the sampling (a warm trap method). The temperature elevation required depends on the RHth characteristic for the specific sorbent, and RH and the temperature of the sampled gas. Usually the 5-15 degreesC elevation is sufficient; only under extreme RH conditions is an elevation of 20 degreesC necessary. The diagrams are given to determine this elevation. Since the sample RH is significantly decreased at an elevated temperature the negative effect of water uptake on the safe sampling volume is alleviated. Consequently the sampled gas volume can be as large as desired which decreases detection limits. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:107 / 116
页数:10
相关论文
共 12 条
[1]   Novel family of multi-layer cartridges filled with a new carbon adsorbent for the quantitative determination of volatile organic compounds in the atmosphere [J].
Brancaleoni, E ;
Scovaventi, M ;
Frattoni, M ;
Mabilia, R ;
Ciccioli, P .
JOURNAL OF CHROMATOGRAPHY A, 1999, 845 (1-2) :317-328
[2]   Stability of reactive low boiling hydrocarbons on carbon based adsorbents typically used for adsorptive enrichment and thermal desorption [J].
Dettmer, K ;
Knobloch, T ;
Engewald, W .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 366 (01) :70-78
[3]   Dry purge for the removal of water from the solid sorbents used to sample volatile organic compounds from the atmospheric air [J].
Gawlowski, J ;
Gierczak, T ;
Pietruszyñska, Z ;
Gawrys, M ;
Niedzielski, J .
ANALYST, 2000, 125 (11) :2112-2117
[4]   Adsorption of water vapour in the solid sorbents used for the sampling of volatile organic compounds [J].
Gawlowski, J ;
Gierczak, T ;
Jezo, A ;
Niedzielski, J .
ANALYST, 1999, 124 (11) :1553-1558
[5]   WATER-ADSORPTION CAPACITY OF THE SOLID ADSORBENTS TENAX-TA, TENAX-GR, CARBOTRAP, CARBOTRAP-C, CARBOSIEVE-SIII, AND CARBOXEN-569 AND WATER MANAGEMENT-TECHNIQUES FOR THE ATMOSPHERIC SAMPLING OF VOLATILE ORGANIC TRACE GASES [J].
HELMIG, D ;
VIERLING, L .
ANALYTICAL CHEMISTRY, 1995, 67 (23) :4380-4386
[6]  
Koch E, 1997, CHEM ANAL-WARSAW, V42, P49
[7]   Water isotherms of activated carbons with small amounts of surface oxygen. [J].
Lodewyckx, P ;
Vansant, EF .
CARBON, 1999, 37 (10) :1647-1649
[8]   Evaluation of a new dualsorbent trap for organic pollutants of different volatility [J].
Sampaolo, L ;
Pierini, E ;
Mastrogiacomo, AR .
CHROMATOGRAPHIA, 1999, 50 (11-12) :680-684
[9]   Recent developments in Dubinin's theory [J].
Stoeckli, F .
CARBON, 1998, 36 (04) :363-368
[10]   WATER-ADSORPTION IN ACTIVE CARBONS DESCRIBED BY THE DUBININ-ASTAKHOV EQUATION [J].
STOECKLI, F ;
JAKUBOV, T ;
LAVANCHY, A .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1994, 90 (05) :783-786