Evaluation of a micro volume pulsed ultrafiltration cell for screening ligands in non-covalent complexes

被引:8
作者
Beverly, MB [1 ]
West, P [1 ]
Julian, RK [1 ]
机构
[1] Eli Lilly & Co, Lilly Corp Ctr, Indianapolis, IN 46285 USA
关键词
D O I
10.2174/1386207023330606
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A pulsed ultrafiltration cell with a 35 muL binding chamber was evaluated for its ability to screen ligands that formed non-covalent complexes with protein targets. The cell was tested with ligands to the targets of carbonic anhydrase and serum albumin. Non-covalent ligand binding to both of these targets was observed and bound ligands were eluted from the cell in less than five min. The cell was also demonstrated to effectively screen a methanolic fermentation broth extract spiked with a known inhibitor to carbonic anhydrase. In addition to detecting specific binding events, the pulsed ultrafiltration method was investigated for its ability to distinguish non-specific binding events. Using carbonic anhydrase with the zinc-binding site removed, it was found that non-specific complexes observed when using electrospray ionization alone were not detected when using the pulsed ultrafiltration mass spectrometry method.
引用
收藏
页码:65 / 73
页数:9
相关论文
共 41 条
  • [1] DOES THE OBSERVATION OF NONCOVALENT COMPLEXES BETWEEN BIOMOLECULES BY ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY NECESSARILY REFLECT SPECIFIC SOLUTION INTERACTIONS
    APLIN, RT
    ROBINSON, CV
    SCHOFIELD, CJ
    WESTWOOD, NJ
    [J]. JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1994, (20) : 2415 - 2417
  • [2] PROBE IMMOBILIZED AFFINITY-CHROMATOGRAPHY MASS-SPECTROMETRY
    BROCKMAN, AH
    ORLANDO, R
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (24) : 4581 - 4585
  • [3] Pulsed ultrafiltration characterization of binding - I. A new method
    Chen, CJ
    Chen, S
    Woodbury, CP
    Venton, DL
    [J]. ANALYTICAL BIOCHEMISTRY, 1998, 261 (02) : 164 - 182
  • [4] USING ELECTROSPRAY-IONIZATION FTICR MASS-SPECTROMETRY TO STUDY COMPETITIVE-BINDING OF INHIBITORS TO CARBONIC-ANHYDRASE
    CHENG, XH
    CHEN, RD
    BRUCE, JE
    SCHWARTZ, BL
    ANDERSON, GA
    HOFSTADLER, SA
    GALE, DC
    SMITH, RD
    GAO, JM
    SIGAL, GB
    MAMMEN, M
    WHITESIDES, GM
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1995, 117 (34) : 8859 - 8860
  • [5] Affinity capillary electrophoresis: From binding measurement to combinatorial library screening
    Chu, YH
    Zang, X
    Tu, J
    [J]. JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 1998, 45 (06) : 713 - 720
  • [6] CONROY CW, 1995, MOL PHARMACOL, V48, P486
  • [7] Iterative size-exclusion chromatography coupled with liquid chromatographic mass spectrometry to enrich and identify tight-binding ligands from complex mixtures
    Davis, RG
    Anderegg, RJ
    Blanchard, SG
    [J]. TETRAHEDRON, 1999, 55 (39) : 11653 - 11667
  • [8] Dunayevskiy YM, 1997, RAPID COMMUN MASS SP, V11, P1178, DOI 10.1002/(SICI)1097-0231(199707)11:11<1178::AID-RCM991>3.0.CO
  • [9] 2-H
  • [10] Affinity-based screening of combinatorial libraries using automated, serial-column chromatography
    Evans, DM
    Williams, KP
    McGuinness, B
    Tarr, G
    Regnier, F
    Afeyan, N
    Jindal, S
    [J]. NATURE BIOTECHNOLOGY, 1996, 14 (04) : 504 - 507