Powered Ankle-Foot Prosthesis Improves Walking Metabolic Economy

被引:368
作者
Au, Samuel K. [1 ]
Weber, Jeff [1 ,2 ]
Herr, Hugh [1 ,3 ]
机构
[1] MIT, Dept Mech Engn, MIT Media Lab, Cambridge, MA 02139 USA
[2] Meka Robot LLC, San Francisco, CA 94107 USA
[3] MIT, Harvard Mit Div Hlth Sci & Technol, Cambridge, MA 02139 USA
关键词
Amputee gait; impedance control; parallel elasticity; powered prosthesis; series elasticity; walking metabolism; KNEE AMPUTEES; WORK; GAIT;
D O I
10.1109/TRO.2008.2008747
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
At moderate to fast walking speeds, the human ankle provides net positive work at high-mechanical-power output to propel the body upward and forward during the stance period. On the contrary, conventional ankle-foot prostheses exhibit a passive-elastic response during stance, and consequently, cannot provide net work. Clinical studies indicate that transtibial amputees using conventional prostheses have higher gait metabolic rates than normal. Researchers believe that the main cause for these higher rates is due to the inability of conventional prostheses to provide sufficient positive power at terminal stance in the trailing leg to limit heel strike losses of the adjacent leading leg. In this investigation, we evaluate the hypothesis that a powered ankle-foot prosthesis, capable of providing human-like ankle work and power during stance, can decrease the metabolic cost of transport (COT) compared to a conventional passive-elastic prosthesis. To test the hypothesis, a powered prosthesis is built that comprises a unidirectional spring, configured in parallel with a force-controllable actuator with series elasticity. The prosthesis is shown to deliver the high mechanical power and net positive work observed in normal human walking. The rate of oxygen consumption and carbon dioxide production is measured as a determinant of metabolic rate on three unilateral transtibial amputees walking at self-selected speeds. We find that the powered prosthesis decreases the amputee's metabolic COT on average by 14% compared to the conventional passive-elastic prostheses evaluated (Flex-Foot Ceterus (R) and Freedom Innovations Sierra), even though the powered system is over twofold heavier than the conventional devices. These results highlight the clinical importance of prosthetic interventions that closely mimic the mass distribution, kinetics, and kinematics of the missing limb.
引用
收藏
页码:51 / 66
页数:16
相关论文
共 51 条
[1]  
[Anonymous], 1992, GAIT ANAL NORMAL PAT
[2]  
Associates H.D, 2001, MEASURE MAN WOMAN HU
[3]  
AU S, 2006, WORKSH DYN WALK MECH
[4]   Powered Ankle-Foot Prosthesis for the improvement of amputee ambulation [J].
Au, Samuel K. ;
Herr, Hugh ;
Weber, Jeff ;
Martinez-Villalpando, Ernesto C. .
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, :3020-3026
[5]  
Au SK, 2008, IEEE ROBOT AUTOM MAG, V15, P52, DOI 10.1109/MRA.2008.927697
[6]   Biomechanical design of a powered ankle-foot prosthesis [J].
Au, Samuel K. ;
Weber, Jeff ;
Herr, Hugh .
2007 IEEE 10TH INTERNATIONAL CONFERENCE ON REHABILITATION ROBOTICS, VOLS 1 AND 2, 2007, :298-303
[7]   An ankle-foot emulation system for the study of human walking biomechanics [J].
Au, Samuel K. ;
Dilworth, Peter ;
Herr, Hugh .
2006 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), VOLS 1-10, 2006, :2939-+
[8]  
Au SK, 2005, INT C REHAB ROBOT, P375
[9]  
AU SK, THESIS MIT CAMBRIDGE
[10]  
BATENI H, 2002, J PROSTHET ORTHOT, V14, P2