Capillary electrochromatography with a silica column with a dynamically modified cationic surfactant

被引:28
作者
Ye, ML [1 ]
Zou, HF [1 ]
Liu, Z [1 ]
Ni, JY [1 ]
Zhang, YK [1 ]
机构
[1] Chinese Acad Sci, Dalian Inst Chem Phys, Natl Chromatog R&A Ctr, Lab Chromatog, Dalian 116011, Peoples R China
基金
中国国家自然科学基金;
关键词
electrochromatography; dynamically modified silica capillary electrochromatography; stationary phases; LC; anilines; benzenes; peptides;
D O I
10.1016/S0021-9673(99)00671-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel mode of capillary electrochromatography (CEC), called dynamically modified silica-capillary electrochromatography, is described in this paper. The column packed with bare silica was dynamically modified with long chain quaternary ammonium salt, cetyltrimethylammonium bromide (CTAB), which was added into the mobile phase. CTAB ions were adsorbed onto the surface of bare silica, and the resulted hydrophobic layer on the silica gel was used as the stationary phase; Using the dynamically modified silica column, neutral solutes were separated by CEC. The highest number of theoretical plates obtained was about 71 500/m and the relative standard deviations for t(0) and capacity factor of toluene were 4.7% and 4.9% for 20 consecutive runs, respectively. The separation mechanism of neutral solutes and the influence of mobile phase composition on the separation was investigated. The separation of nitrogen-containing solutes was carried out with this mode and the peak tailing of basic solute was effectively eliminated because the adsorption of basic solute on silica was blocked by the preferred adsorption of CTAB. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 23 条
[1]   A review of the current status of capillary electrochromatography technology and applications [J].
Altria, KD ;
Smith, NW ;
Turnbull, CH .
CHROMATOGRAPHIA, 1997, 46 (11-12) :664-674
[2]   PARTITIONING BEHAVIOR OF SOLUTES ELUTED WITH MICELLAR MOBILE PHASES IN LIQUID-CHROMATOGRAPHY [J].
ARMSTRONG, DW ;
NOME, F .
ANALYTICAL CHEMISTRY, 1981, 53 (11) :1662-1666
[3]   PRESSURIZED GRADIENT ELECTRO-HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BEHNKE, B ;
BAYER, E .
JOURNAL OF CHROMATOGRAPHY A, 1994, 680 (01) :93-98
[4]   SURFACE SILANOLS IN SILICA-BONDED HYDROCARBONACEOUS STATIONARY PHASES .2. IRREGULAR RETENTION BEHAVIOR AND EFFECT OF SILANOL MASKING [J].
BIJ, KE ;
HORVATH, C ;
MELANDER, WR ;
NAHUM, A .
JOURNAL OF CHROMATOGRAPHY, 1981, 203 (JAN) :65-84
[5]   Electrochromatography [J].
Crego, AL ;
Gonzalez, A ;
Marina, ML .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1996, 26 (04) :261-304
[6]   IMPROVED CAPILLARY ZONE ELECTROPHORETIC SEPARATION OF BASIC-PROTEINS, USING A FLUOROSURFACTANT BUFFER ADDITIVE [J].
EMMER, A ;
JANSSON, M ;
ROERAADE, J .
JOURNAL OF CHROMATOGRAPHY, 1991, 547 (1-2) :544-550
[7]   HYDROPHOBIC CHROMATOGRAPHY WITH DYNAMICALLY COATED STATIONARY PHASES [J].
GHAEMI, Y ;
WALL, RA .
JOURNAL OF CHROMATOGRAPHY, 1979, 174 (01) :51-59
[8]   Separation of pharmaceutical bases from neutral and acidic components by capillary electrochromatography [J].
Gillott, NC ;
Euerby, MR ;
Johnson, CM ;
Barrett, DA ;
Shaw, PN .
ANALYTICAL COMMUNICATIONS, 1998, 35 (07) :217-220
[9]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON DYNAMICALLY MODIFIED SILICA .3. MODIFICATION OF SILICA WITH LONG-CHAIN AND SYMMETRICAL QUATERNARY AMMONIUM-COMPOUNDS [J].
HANSEN, SH ;
HELBOE, P ;
LUND, U .
JOURNAL OF CHROMATOGRAPHY, 1982, 240 (02) :319-327
[10]  
HELBOE P, 1989, ADV CHROMATOGR, V28, P195