Fiber-optic evanescent-wave spectroscopy for fast multicomponent analysis of human blood

被引:35
作者
Simhi, R [1 ]
Gotshal, Y [1 ]
Bunimovich, D [1 ]
Sela, BA [1 ]
Katzir, A [1 ]
机构
[1] CHAIM SHEBA MED CTR,INST CHEM PATHOL,IL-52621 TEL HASHOMER,ISRAEL
来源
APPLIED OPTICS | 1996年 / 35卷 / 19期
关键词
attenuated total-reflection spectroscopy; fiber-optic evanescent-wave spectroscopy; Fourier-transform infrared spectrometer; blood; multivariate calibration; neural networks;
D O I
10.1364/AO.35.003421
中图分类号
O43 [光学];
学科分类号
070207 [光学]; 0803 [光学工程];
摘要
A spectral analysis of human blood serum was undertaken by fiber-optic evanescent-wave spectroscopy (FEWS) by the use of a Fourier-transform infrared spectrometer. A special cell for the FEWS measurements was designed and built that incorporates an IR-transmitting silver halide fiber and a means for introducing the blood-serum sample. Further improvements in analysis were obtained by the adoption of multivariate calibration techniques that are already used in clinical chemistry. The partial least-squares algorithm was used to calculate the concentrations of cholesterol, total protein, urea, and uric acid in human blood serum. The estimated prediction errors obtained (in percent from the average value) were 6% for total protein, 15% for cholesterol, 30% for urea, and 30% for uric acid. These results were compared with another independent prediction method that used a neural-network model. This model yielded estimated prediction errors of 8.8% for total protein, 25% for cholesterol, and 21% for uric acid. (C) 1996 Optical Society ofAmerica
引用
收藏
页码:3421 / 3425
页数:5
相关论文
共 16 条
[1]
GLUCOSE DETERMINATION IN SIMULATED BLOOD-SERUM SOLUTIONS BY FOURIER-TRANSFORM INFRARED-SPECTROSCOPY - INVESTIGATION OF SPECTRAL INTERFERENCES [J].
BHANDARE, P ;
MENDELSON, Y ;
STOHR, E ;
PEURA, RA .
VIBRATIONAL SPECTROSCOPY, 1994, 6 (03) :363-378
[2]
Born M., 1986, PRINCIPLES OPTICS
[3]
A SYSTEM FOR MONITORING AND CONTROL OF PROCESSES BASED ON IR FIBERS AND TUNABLE DIODE-LASERS [J].
BUNIMOVICH, D ;
KELLNER, R ;
KRSKA, R ;
MESICA, A ;
PAISS, I ;
SCHIESL, U ;
TACKE, M ;
TAGA, K ;
KATZIR, A .
JOURNAL OF MOLECULAR STRUCTURE, 1993, 292 :125-132
[4]
Harrick N.J., 1979, INTERNAL REFLECTION
[5]
MULTIVARIATE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD BY ATTENUATED TOTAL REFLECTION INFRARED-SPECTROSCOPY [J].
HEISE, HM ;
MARBACH, R ;
JANATSCH, G ;
KRUSEJARRES, JD .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2009-2015
[6]
MULTICOMPONENT ASSAY FOR BLOOD SUBSTRATES IN HUMAN PLASMA BY MIDINFRARED SPECTROSCOPY AND ITS EVALUATION FOR CLINICAL ANALYSIS [J].
HEISE, HM ;
MARBACH, R ;
KOSCHINSKY, T ;
GRIES, FA .
APPLIED SPECTROSCOPY, 1994, 48 (01) :85-95
[7]
Henry T.J., 1974, CLIN CHEM PRINCIPLES
[8]
HERTZ J, 1991, INTRO THEORY NEURAL, pCH1
[9]
MULTIVARIATE CALIBRATION FOR ASSAYS IN CLINICAL-CHEMISTRY USING ATTENUATED TOTAL REFLECTION INFRARED-SPECTRA OF HUMAN-BLOOD PLASMA [J].
JANATSCH, G ;
KRUSEJARRES, JD ;
MARBACH, R ;
HEISE, HM .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2016-2023
[10]
KATZ M, 1994, APPL OPTICS, V33, P588