Detection of swallows with silent aspiration using swallowing and breath sound analysis

被引:25
作者
Shirazi, Samaneh Sarraf
Buchel, Caitlin [2 ]
Daun, Reesa [2 ]
Lenton, Laura [2 ]
Moussavi, Zahra [1 ]
机构
[1] Univ Manitoba, Canada Res Chair Biomed Engn, Dept Elect & Comp Engn, Winnipeg, MB R3T 5V6, Canada
[2] Riverview Hlth Ctr, Dept Speech Language Therapy, Winnipeg, MB R3L 2P4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Acoustical analysis; Breath sound analysis; Silent aspiration; Swallowing; Clustering; CERVICAL AUSCULTATION; DYSPHAGIA; STROKE; RELIABILITY; VALIDITY; AIRWAY; RISK; CLASSIFICATION; DISORDERS; FLOW;
D O I
10.1007/s11517-012-0958-9
中图分类号
TP39 [计算机的应用];
学科分类号
080201 [机械制造及其自动化];
摘要
In this study, the feasibility of acoustical analysis for detection of swallowing silent aspiration is investigated. As a pilot study, we analyzed the breath sounds of 21 dysphagic individuals, 11 of which demonstrated aspiration during the fiberoptic endoscopic evaluation of swallowing (FEES) or videofluoroscopic swallowing study (VFSS). We found that the low frequency components of the power spectrum of the breath sounds after a swallow show higher magnitude when there is aspiration. Thus, we divided the frequency range below 300 Hz into three sub-bands and calculated the average power of the breath sound signal in each sub-band as the characteristic features for the stage 1 classification into two groups of aspirated and non-aspirated patients. Then, for the aspirated group, the unsupervised fuzzy k-means clustering algorithm was deployed to label the breath sounds immediately after a swallow as aspiration or non-aspiration. The results were compared with the FEES/VFSS assessments provided by the speech language pathologists. The results are encouraging: more than 86 % accuracy in detection of silent aspiration. While the proposed method should be verified on a larger dataset, the results are promising for the use of acoustical analysis as a clinical tool to detect silent aspiration.
引用
收藏
页码:1261 / 1268
页数:8
相关论文
共 32 条
[1]
Aboofazeli M, 2005, INT CONF ACOUST SPEE, P421
[2]
Aboofazeli Mohammad, 2006, Conf Proc IEEE Eng Med Biol Soc, V2006, P5591
[3]
[Anonymous], 1998, EVALUATION TREATMENT
[4]
The upper airway in sleep: physiology of the pharynx [J].
Ayappa, I ;
Rapoport, DM .
SLEEP MEDICINE REVIEWS, 2003, 7 (01) :9-33
[5]
Bezdek J.C., 1981, PATTERN RECOGNITION, P272
[6]
Blevins RD., 1990, Flow-induced vibrations
[7]
Reliability and validity of cervical auscultation [J].
Borr, Christiane ;
Hielscher-Fastabend, Martina ;
Luecking, Andy .
DYSPHAGIA, 2007, 22 (03) :225-234
[8]
Dempsey J. E., 1990, S DYSPB J HOPK HOSP
[9]
Duda R.O., 2000, PATTERN CLASSIFICATI, V2, P654
[10]
HAMLET S L, 1992, Dysphagia, V7, P160, DOI 10.1007/BF02493450