Influence of high-energy photons on the spectrum of iodine-123 with low- and medium-energy collimators:: consequences for imaging with 123I-labelled compounds in clinical practice

被引:67
作者
Dobbeleir, AA
Hambye, ASE
Franken, PR
机构
[1] Middelheim Hosp, B-2020 Antwerp, Belgium
[2] Free Univ Brussels, VUB, Brussels, Belgium
关键词
iodine-123; high energy photons; spectral analysis; collimator;
D O I
10.1007/s002590050434
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Using iodine-123 labelled radiotracers, the presence of 2.5% high-energy photons causes image deterioration due to increased scatter. To investigate the influence of these photons on image quality, we measured the spectrum of I-123 With a medium-energy (ME), a low-energy all-purpose (LEAP) and a low-energy high-resolution (LEHR) collimator. Even in air, using low-energy collimators a high baseline activity was observed over the total energy detection range of the gamma camera. The 159-keV photopeak to scatter activity ratio fell from 5.9 for ME to 3.6 and 2.9 for LE collimators. Acquisition of images with LEHR collimators with energy windows set at 159 keV and 500 keV demonstrated that the 159-keV LEHR image is a combination of the ME image of the object and of the LEHR 500-keV image. Because of their important septal penetration and greater geometric detection efficiency compared with the 159-keV photons of I-123, the contribution of high-energy photons is dependent on the source-detector distance. For a small source placed in air, the scatter to photopeak activities varied from 17.4% at 80 cm to 37.8% at 5 cm distance from an LEHR collimator. Considering only the scatter problem, ME collimators are the best choice for I-123 studies. When using LE collimators for high-resolution tomography with I-123-labelled compounds, scatter contribution from high-energy photons has to be corrected for quantitative analysis or when dual-isotope studies are performed, whether or not these studies are acquired simultaneously.
引用
收藏
页码:655 / 658
页数:4
相关论文
共 19 条
[1]   Optimal collimator choice for sequential iodine-123 and technetium-99m imaging [J].
DeGeeter, F ;
Franken, PR ;
Defrise, M ;
Andries, H ;
Saelens, E ;
Bossuyt, A .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1996, 23 (07) :768-774
[2]  
DOBBELEIR A, 1999, J NUCL MED, V40
[3]  
Franken PR, 1996, J NUCL MED, V37, P718
[4]  
GILLAND DR, 1994, J NUCL MED, V35, P1707
[5]   Recovery of impaired left ventricular function in patients with acute myocardial infarction is predicted by the discordance in defect size on I-123-BMIPP and Tl-201 SPET images [J].
Ito, T ;
Tanouchi, J ;
Kato, J ;
Morioka, T ;
Nishino, M ;
Iwai, K ;
Tanahashi, H ;
Yamada, Y ;
Hori, M ;
Kamada, T .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1996, 23 (08) :917-923
[6]   Use of 123I-BMIPP single-photon emission tomography to estimate areas at risk following successful revascularization in patients with acute myocardial infarction [J].
Kawai, Y ;
Tsukamoto, E ;
Nozaki, Y ;
Kishino, K ;
Kohya, T ;
Tamaki, N .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (10) :1390-1395
[7]  
Kobayashi H, 1998, J NUCL MED, V39, P1117
[8]  
MACEY DJ, 1986, J NUCL MED, V27, P1467
[9]   Differential distribution of striatal [123l]β-CIT in Parkinson's disease and progressive supranuclear palsy, evaluated with single-photon emission tomography [J].
Messa, C ;
Volonté, MA ;
Fazio, F ;
Zito, F ;
Carpinelli, A ;
d'Amico, A ;
Rizzo, G ;
Moresco, RM ;
Paulesu, E ;
Franceschi, M ;
Lucignani, G .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (09) :1270-1276
[10]  
Nakata T, 1998, J NUCL MED, V39, P1492