Accuracy of a video strain measurement system

被引:21
作者
Smutz, WP [1 ]
Drexler, M [1 ]
Berglund, LJ [1 ]
Growney, E [1 ]
An, KN [1 ]
机构
[1] MAYO CLIN & MAYO FDN,ORTHOPED BIOMECH LAB,ROCHESTER,MN 55905
关键词
strain measurement; soft tissue; video;
D O I
10.1016/0021-9290(95)00131-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The purpose of this study was to determine the accuracy of a video system which our laboratory has been using to measure soft tissue strain. Both static and dynamic error analyses were performed to assess the accuracy of our video system. Static error was defined as the amount of movement reported by the video system for markers that were stationary. Dynamic error was defined as the difference between the motion of the markers as reported by the video system and their actual motion. Two sets of fluorescent markers were attached to a servo-hydraulic materials test machine. One marker set was attached to the hydraulic actuator (moving markers) and the other set was attached to the base of the machine (stationary markers). Five different marker sizes, Eve camera distances, and seven different loading rates were studied. Results indicated that the static error was independent of marker size, and that the dynamic error was independent of the loading rate and marker size for loading rates of 50% of the camera field of view (CFV) per second or slower. For loading rates greater than 50% of CFV per second, the marker size did have an affect on the dynamic error. The mean static error was found to be 0.026% of CFV and the mean dynamic error was found to be 0.062% of CFV. (C) 1996 Elsevier Science Ltd.
引用
收藏
页码:813 / 817
页数:5
相关论文
共 11 条
[1]   STRAINS IN THE ARTICULAR DISK OF THE TRIANGULAR FIBROCARTILAGE COMPLEX - A BIOMECHANICAL STUDY [J].
ADAMS, BD ;
HOLLEY, KA .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1993, 18A (05) :919-925
[2]   BIOCHEMICALLY DISCRETE ZONES OF CANINE FLEXOR TENDON - EVALUATION OF PROPERTIES WITH A NEW PHOTOGRAPHIC METHOD [J].
AMADIO, PC ;
BERGLUND, LJ ;
AN, KN .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1992, 10 (02) :198-204
[3]   EFFECTS OF STRUCTURE AND STRAIN-MEASUREMENT TECHNIQUE ON THE MATERIAL PROPERTIES OF YOUNG HUMAN TENDONS AND FASCIA [J].
BUTLER, DL ;
GROOD, ES ;
NOYES, FR ;
ZERNICKE, RF ;
BRACKETT, K .
JOURNAL OF BIOMECHANICS, 1984, 17 (08) :579-596
[4]   A NEW OPTICAL-SYSTEM FOR THE DETERMINATION OF DEFORMATIONS AND STRAINS - CALIBRATION CHARACTERISTICS AND EXPERIMENTAL RESULTS [J].
DERWIN, KA ;
SOSLOWSKY, LJ ;
GREEN, WDK ;
ELDER, SH .
JOURNAL OF BIOMECHANICS, 1994, 27 (10) :1277-+
[5]   AN IMPROVED VIDEO-BASED COMPUTER TRACKING SYSTEM FOR SOFT BIOMATERIALS TESTING [J].
DOWNS, J ;
HALPERIN, HR ;
HUMPHREY, J ;
YIN, F .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (09) :903-907
[6]   ON-LINE MEASUREMENT OF MICROVASCULAR DIMENSIONS BY TELEVISION MICROSCOPY [J].
INTAGLIETTA, M ;
TOMPKINS, WR .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 32 (04) :546-+
[7]   CALIBRATION CHARACTERISTICS OF A VIDEO DIMENSION ANALYZER (VDA) SYSTEM [J].
LAM, TC ;
FRANK, CB ;
SHRIVE, NG .
JOURNAL OF BIOMECHANICS, 1992, 25 (10) :1227-1231
[8]   2-DIMENSIONAL MECHANICAL-PROPERTIES OF RABBIT SKIN .1. EXPERIMENTAL SYSTEM [J].
LANIR, Y ;
FUNG, YC .
JOURNAL OF BIOMECHANICS, 1974, 7 (01) :29-34
[9]   AGING, VERTEBRAL DENSITY, AND DISC DEGENERATION ALTER THE TENSILE STRESS-STRAIN CHARACTERISTICS OF THE HUMAN ANTERIOR LONGITUDINAL LIGAMENT [J].
NEUMANN, P ;
EKSTROM, LA ;
KELLER, TS ;
PERRY, L ;
HANSSON, TH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1994, 12 (01) :103-112
[10]   MEASUREMENTS OF NONHOMOGENEOUS, DIRECTIONAL MECHANICAL-PROPERTIES OF ARTICULAR-CARTILAGE IN TENSION [J].
WOO, SLY ;
AKESON, WH ;
JEMMOTT, GF .
JOURNAL OF BIOMECHANICS, 1976, 9 (12) :785-791