共 4 条
Design and performance of a stable first crystal mount for a cryogenically cooled Si monochromator at the Advanced Photon Source
被引:10
作者:
Dufresne, EM
[1
]
Arms, DA
Dierker, SB
Clarke, R
Yacoby, Y
Pitney, J
MacHarrie, B
Pindak, R
机构:
[1] Univ Michigan, Dept Phys, Ann Arbor, MI 48109 USA
[2] Hebrew Univ Jerusalem, Racah Inst Phys, IL-91904 Jerusalem, Israel
[3] Lucent Technol, Bell Labs, Murray Hill, NJ 07974 USA
[4] Brookhaven Natl Lab, Natl Synchrotron Light Source, Upton, NY 11973 USA
关键词:
D O I:
10.1063/1.1423630
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
We present a new design for mounting a cryogenically cooled Si crystal which gives greatly improved beam stability. The design has been successfully implemented at the University of Michigan, Howard University, Bell Laboratories-Lucent Technologies Collaborative Access Team (MHATT-CAT) 7ID Beamline of the Advanced Photon Source. Before the installation of the new crystal mount, our Si (111) cryogenically cooled monochromator was sensitive to the pressure fluctuations of the liquid nitrogen coolant, such that the angle of incidence on the first crystal varied linearly with the applied pressure in the cooling lines, causing beam motion of about 250 mum, 60 m from the source. The key element of the design is a symmetrically positioned cooling manifold which balances the forces caused by pressure fluctuations. With this new mount, the typical beam stability is now about 10 mum, comparable to the source stability. (C) 2002 American Institute of Physics.
引用
收藏
页码:1511 / 1513
页数:3
相关论文