Monte Carlo investigation of electron beam output factors versus size of square cutout

被引:54
作者
Zhang, GG [2 ]
Rogers, DWO
Cygler, JE
Mackie, TR
机构
[1] NRC, Ottawa, ON K1S 5B6, Canada
[2] Carleton Univ, Ottawa, ON K1S 5B6, Canada
[3] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
[4] Ottawa Reg Canc Ctr, Ottawa, ON, Canada
[5] Univ Wisconsin, Madison, WI USA
关键词
electron beam dosimetry; relative output factors; cutout factors; Monte Carlo;
D O I
10.1118/1.598582
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A major task in commissioning an electron accelerator is to measure relative output factors versus cutout size (i.e., cutout factors) for various electron beam energies and applicator sizes. We use the BEAM Monte Carlo code [Med Phys. 22, 503-524 (1995)] to stimulate clinical electron beams and to calculate the relative output factors for square cutouts. Calculations are performed for a Siemens MD2 linear accelerator with beam energies, 6, 9, 11, and 13 MeV. The calculated cutout factors for square cutouts in 10X10 cm(2), 15X15 cm(2), and 20X20 cm(2) applicators at SSDs of 100 and 115 cm agree with the measurements made using a silicon diode within about 1% except for the smallest cutouts at SSD=115 cm where they agree within 0.015. The details of each component of the dose, such as the dose from particles scattered off the jaws and the applicator, the dose from contaminant photons, the dose from direct electrons, etc., are also analyzed. The calculations show that in-phantom side-scatter equilibrium is a major factor for the contribution from the direct component which usually dominates the output of a beam. It takes about 6 h of CPU time on a Pentium Pro 200MHz computer to simulate an accelerator and additional 2 h to calculate the relative output factor for each cutout with a statistical uncertainty of 1%. [S0094-2405(99)01405-4].
引用
收藏
页码:743 / 750
页数:8
相关论文
共 13 条
[1]   CALCULATION OF ELECTRON-BEAM DEPTH-DOSE CURVES AND OUTPUT FACTORS FOR ARBITRARY FIELD SHAPES [J].
BRUINVIS, IAD ;
MATHOL, WAF .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (04) :395-404
[2]  
BRUN R, 1992, PAW USER GUIDE
[3]   Measurements of electron beam peak scatter factors [J].
Ding, GX ;
Cygler, JE .
MEDICAL PHYSICS, 1998, 25 (02) :251-253
[4]   CALCULATION OF STOPPING POWER RATIOS USING REALISTIC CLINICAL ELECTRON-BEAMS [J].
DING, GX ;
ROGERS, DWO ;
MACKIE, TR .
MEDICAL PHYSICS, 1995, 22 (05) :489-501
[5]   CLINICAL ELECTRON-BEAM DOSIMETRY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-25 [J].
KHAN, FM ;
DOPPKE, KP ;
HOGSTROM, KR ;
KUTCHER, GJ ;
NATH, R ;
PRASAD, SC ;
PURDY, JA ;
ROZENFELD, M ;
WERNER, BL .
MEDICAL PHYSICS, 1991, 18 (01) :73-109
[6]   A METHOD OF CALCULATING THE OUTPUT FACTORS OF ARBITRARILY SHAPED ELECTRON FIELDS [J].
MCPARLAND, BJ .
MEDICAL PHYSICS, 1989, 16 (01) :88-93
[7]   PREDICTION OF ELECTRON-BEAM OUTPUT FACTORS [J].
MILLS, MD ;
HOGSTROM, KR ;
ALMOND, PR .
MEDICAL PHYSICS, 1982, 9 (01) :60-68
[8]   DETERMINATION OF ELECTRON-BEAM OUTPUT FACTORS FOR A 20-MEV LINEAR-ACCELERATOR [J].
MILLS, MD ;
HOGSTROM, KR ;
FIELDS, RS .
MEDICAL PHYSICS, 1985, 12 (04) :473-476
[9]  
Nelson W., 1985, SLAC265
[10]  
NELSON WR, 1989, MONTE CARLO TRANSPOR, P287