Determination of the noise source in the electrocardiogram during cardiopulmonary resuscitation

被引:34
作者
Fitzgibbon, E
Berger, R
Tsitlik, J
Halperin, HR
机构
[1] Johns Hopkins Univ, Sch Med, Dept Med, Div Cardiol, Baltimore, MD 21287 USA
[2] Johns Hopkins Univ, Sch Med, Dept Radiol, Baltimore, MD 21287 USA
[3] Johns Hopkins Univ, Sch Med, Dept Biomed Engn, Baltimore, MD 21287 USA
关键词
electrocardiogram; cardiopulmonary resuscitation; electrode; noise;
D O I
10.1097/00003246-200204001-00006
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
During cardiopulmonary resuscitation (CPR), the electrocardiogram (ECG) is often obscured by noise. This noise is in the form of baseline variations in the EGG, which often necessitate stopping chest compressions to adequately assess the ECG. Because survival from cardiac arrest has been shown to be related to blood flow generated during CPR, and because interruption of chest compressions will reduce blood flow, survival may be compromised by these interruptions. Three possible sources for the noise were considered: the heart, which is deformed during CPR, which may introduce a mechanical-electrical interaction and alter the normal electrical pattern of the heart; the thoracic cavity, which may have large impedance variations because of CPR and thereby modulate the ECG; and the skin-electrode interface, which may be mechanically disturbed during CPR and thus produce polarization potentials that cause additional noise. CPR studies were performed on five dogs by using four different test conditions and six different types of electrodes. The test conditions were: electrode motion, which allowed mechanical disturbances of the skin-electrode interface without altering the thoracic impedance or deforming the heart; vest CPR; manual CPR; and respiration. The myocardial EGG, the bipolar and unipolar surface ECGs, and the thoracic impedance were monitored. Different types of surface EGG electrodes were used to determine whether the noise was dependent on electrode type or size. There were no baseline variations in the myocardial ECG during any of the test conditions. The thoracic impedance did vary during CPR, but the variations were temporally uncorrelated to the baseline variations in the ECG, and the variations were of similar magnitude as the variations caused by respiration, which produced no baseline changes in the ECG. Finally, the magnitude of the baseline variations In the ECG was substantially different for electrodes of different sizes and shapes, and electrode motion produced baseline variations that were identical to those produced during CPR. Therefore, it was concluded that the source of the noise in the EGG during CPR is the skin-electrode interface and, specifically, that the noise is related to the electrical properties of the electrode.
引用
收藏
页码:S148 / S153
页数:6
相关论文
共 44 条
[1]  
*AM HEART ASS, 1988, JAMA-J AM MED ASSOC, V255, P2841
[2]  
*AM HEART ASS, 1983, TXB ADV CARD LIF SUP
[3]  
CHU CH, 1991, J ELECTROCARDIOL, V23, P192
[4]   IMPULSIVE NOISE SUPPRESSION AND BACKGROUND NORMALIZATION OF ELECTROCARDIOGRAM SIGNALS USING MORPHOLOGICAL OPERATORS [J].
CHU, CHH ;
DELP, EJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (02) :262-273
[5]  
Dahl C F, 1976, Med Instrum, V10, P151
[6]   ELECTRODE PAD SIZE, TRANS-THORACIC IMPEDANCE AND SUCCESS OF EXTERNAL VENTRICULAR DEFIBRILLATION [J].
DALZELL, GWN ;
CUNNINGHAM, SR ;
ANDERSON, J ;
ADGEY, AAJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (12) :741-744
[7]  
DOWNS J, 1991, ECG AMPLIFIER REPORT
[8]   CANINE TRANSTHORACIC RESISTANCE [J].
EWY, GA ;
EWY, MD ;
NUTTALL, AW ;
NUTTALL, AJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 32 (01) :91-&
[9]   PROBLEMS AND LIMITATIONS OF ECG BASELINE ESTIMATION AND REMOVAL USING A CUBIC SPLINE TECHNIQUE DURING EXERCISE ECG TESTING - RECOMMENDATIONS FOR PROPER IMPLEMENTATION [J].
FRONING, JN ;
OLSON, MD ;
FROELICHER, VF .
JOURNAL OF ELECTROCARDIOLOGY, 1988, 21 :S149-S157
[10]  
Geddes L A, 1976, Med Instrum, V10, P159