Screening applications for MRI in the detection of upper abdominal disease: comparative study of non-contrast-enhanced single-shot MRI and contrast-enhanced helical CT

被引:8
作者
De Jaegere, T [1 ]
Van Hoe, L [1 ]
Van Steenbergen, W [1 ]
Van Cutsem, E [1 ]
Bosmans, H [1 ]
Heindryckx, E [1 ]
Loubeyre, P [1 ]
Marchal, G [1 ]
机构
[1] Univ Hosp Gasthuisberg, Dept Radiol, B-3000 Louvain, Belgium
关键词
magnetic resonance imaging; rapid imaging; magnetic resonance; comparative studies; computed tomography; helial technology;
D O I
10.1007/s003300050755
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the value of 'push-button' singe-shot non-contrast-enhanced MRI and contrast-enhanced helical CT for detection of upper abdominal disease. Methods: In 120 patients, images obtained with non contrast-enhanced single-shot MRI (T2: double echo HASTE, and T1: turbo, FLASH) and contrast-enhanced helical CT were compared. Lesions or abnormalities were divided in 8 anatomical categories (1: liver; 2: pancreatobiliary; 3: kidney/adrenal gland; 4: retroperitoneum; 5: vascular; 6: spleen; 7: gastrointestinal tract and peritoneum; 8: base of thorax) and classified as follows: 2: seen at !;MRI only; 1: better seen at MRT; 0: no difference; -1: better seen at CT; -2: seen at CT only. Also recorded were the 'door-to-door' examination times. Results! Of a total of 629 abnormalities, 594 were detected at MRI (94 %) and 536 at CT (85 %). CT offered better results in two categories only: retroperitoneum (mean score: -0.68) and vascular (mean score -0.87). Mean examination times were 19 min for CT and 14.8 min for MRI. Conclusion: Unenhanced single-shot MRI is a valuable first step of a comprehensive upper abdominal MR exam and may: even be the final step in many patients.
引用
收藏
页码:853 / 861
页数:9
相关论文
共 21 条
[1]   Breath-hold MR urography using the HASTE technique [J].
Aerts, P ;
VanHoe, L ;
Bosmans, H ;
Oyen, R ;
Marchal, G ;
Baert, AL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (03) :543-545
[2]  
BELL RA, 1996, JMRI-J MAGN RESON IM, V1, P10
[3]   Single-shot T2-weighted MR imaging of the upper abdomen: Preliminary experience with double-echo HASTE technique [J].
Bosmans, H ;
VanHoe, L ;
Gryspeerdt, S ;
Kiefer, B ;
VanSteenbergen, W ;
Baert, AL ;
Marchal, G .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 169 (05) :1291-1293
[4]   Preliminary experience with a new double-echo half-Fourier single-shot turbo spin echo acquisition in the characterization of liver lesions [J].
Bosmans, H ;
Gryspeerdt, S ;
VanHoe, L ;
VanOostende, S ;
DeJaegere, T ;
Kiefer, B ;
Baert, AL ;
Marchal, G .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 1997, 5 (01) :79-84
[5]  
Gaa J, 1996, EUR RADIOL, V6, P838
[6]   SPIRAL(HELICAL) CT [J].
HEIKEN, JP ;
BRINK, JA ;
VANNIER, MW .
RADIOLOGY, 1993, 189 (03) :647-656
[7]   T1-WEIGHTED SNAPSHOT GRADIENT-ECHO MR IMAGING OF THE ABDOMEN [J].
HOLSINGERBAMPTON, AE ;
RIEDERER, SJ ;
CAMPEAU, NG ;
EHMAN, RL ;
JOHNSON, CD .
RADIOLOGY, 1991, 181 (01) :25-32
[8]   Diagnostic value of MR imaging in comparison to CT in the detection and differential diagnosis of renal masses: ROC analysis [J].
Kreft, BP ;
MullerMiny, H ;
Sommer, T ;
Steudel, A ;
Vahlensieck, M ;
Novak, D ;
Muller, BG ;
Schild, HH .
EUROPEAN RADIOLOGY, 1997, 7 (04) :542-547
[9]   Two-phase helical CT for pancreatic tumors: Pancreatic versus hepatic phase enhancement of tumor, pancreas, and vascular structures [J].
Lu, DSK ;
Vedantham, S ;
Krasny, RM ;
Kadell, B ;
Berger, WL ;
Reber, HA .
RADIOLOGY, 1996, 199 (03) :697-701
[10]   Fast MR imaging techniques: Impact in the abdomen [J].
Mitchell, DG .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1996, 6 (05) :812-821