Scanner and kVp dependence of measured CT numbers in the ACR CT phantom

被引:105
作者
Cropp, Robert J. [1 ]
Seslija, Petar [1 ]
Tso, David [2 ]
Thakur, Yogesh [1 ,2 ]
机构
[1] Vancouver Coastal Hlth, Integrated Med Imaging, Vancouver, BC, Canada
[2] Univ British Columbia, Fac Med, Dept Radiol, Vancouver, BC, Canada
关键词
CT number; ACR phantom; CT quality control; COMPUTED-TOMOGRAPHY; STONE COMPOSITION; CALCIUM; RADIATION; SIZE;
D O I
10.1120/jacmp.v14i6.4417
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Quality control testing of CT scanners in our region includes a measurement of CT numbers in the American College of Radiology (ACR) CT phantom using a standardized protocol. CT number values are clinically relevant in determining the composition of various tissues in the body. Accuracy is important in the characterization of tumors, assessment of coronary calcium, and identification of urinary stone composition. Effective quality control requires that tolerance ranges of CT number values be defined: a measured value outside the range indicates the need for further investigation and possible recalibration of the scanner. This paper presents the results of CT number measurements on 36 scanners (25 GE, 10 Siemens and 1 Toshiba) at each available kVp. Among the five materials (solid water, air, polyethylene, acrylic, bone-equivalent) the measured CT numbers exhibit manufacturer and kVp dependence, which should be taken into account when defining tolerances. With this scan protocol, air and solid water values are significantly higher on GE scanners than on Siemens scanners (p-value < 0.01 at each kVp). The CT numbers of polyethylene and acrylic increase with kVp, while the bone-equivalent CT number decreases. These results are used to define manufacturer-and kVp-specific tolerance ranges for the CT numbers of each material in this phantom, which will be used in our quality control program.
引用
收藏
页码:338 / 349
页数:12
相关论文
共 21 条
[1]
ACR, 2012, CT ACCR PHANT PROGR
[2]
QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY [J].
AGATSTON, AS ;
JANOWITZ, WR ;
HILDNER, FJ ;
ZUSMER, NR ;
VIAMONTE, M ;
DETRANO, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :827-832
[3]
Adrenal masses: Characterization with combined unenhanced and delayed enhanced CT [J].
Caoili, EM ;
Korobkin, M ;
Francis, IR ;
Cohan, RH ;
Platt, JF ;
Dunnick, NR ;
Raghupathi, KI .
RADIOLOGY, 2002, 222 (03) :629-633
[4]
Diagnosis of Renal Angiomyolipoma with Hounsfield Unit Thresholds: Effect of Size of Region of Interest and Nephrographic Phase Imaging [J].
Davenport, Matthew S. ;
Neville, Amy M. ;
Ellis, James H. ;
Cohan, Richard H. ;
Chaudhry, Humaira S. ;
Leder, Richard A. .
RADIOLOGY, 2011, 260 (01) :158-165
[5]
European Commission, 1999, 16262 EUR EU
[6]
First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[7]
From the RSNA refresher courses - A practical approach to the cystic renal mass [J].
Hartman, DS ;
Choyke, PL ;
Hartman, MS .
RADIOGRAPHICS, 2004, 24 :S101-S115
[8]
Health Canada, 2008, Safety Code 35: Safety Procedures for the Installation, Use and Control of X-ray Equipment in Large Medical Radiological Facilities
[9]
Determination of Renal Stone Composition with Dual-Energy CT: In Vivo Analysis and Comparison with X-ray Diffraction [J].
Hidas, Guy ;
Eliahou, Ruth ;
Duvdevani, Mordechai ;
Coulon, Phillipe ;
Lemaitre, Laurent ;
Gofrit, Ofer N. ;
Pode, Dov ;
Sosna, Jacob .
RADIOLOGY, 2010, 257 (02) :394-401
[10]
Age and gender distributions of coronary artery calcium detected by electron beam tomography in 35,246 adults [J].
Hoff, JA ;
Chomka, EV ;
Krainik, AJ ;
Daviglus, M ;
Rich, S ;
Kondos, GT .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 87 (12) :1335-1339