RF heating of implanted spinal fusion stimulator during magnetic resonance imaging

被引:63
作者
Chou, CK
McDougall, JA
Chan, KW
机构
[1] Department of Radiation Research, City of Hope National Medical Center, Duarte, CA 91010-3000
关键词
bone fusion; induced current; phantom model; RF energy absorption; thermal effect;
D O I
10.1109/10.568912
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Radio frequency (RF) heating of an implanted spinal fusion stimulator (SpF) during magnetic resonance imaging (MRI) was studied on a full-size human phantom. Heating during MRI scans (GE Signa 4X, 1.5 T) was measured with RF-transparent fiberoptic sensors. With the implant correctly connected, the maximum temperature rises were less than 2 degrees C during the 26 min that the scans were at maximum RF power. At the tip of a broken stimulator lead (connecting the SpF generator and its electrodes), the maximum temperature rise was 11-14 degrees C. Regular 4-min scans of the spinal cord produced similar temperature rises at the broken tip. After the generator and the leads were removed, heating at the electrode connector tip was less than 1.5 degrees C. The control temperature rises at the same locations, without the stimulator, were less than 0.5 degrees C. This study shows that spinal fusion stimulator heating is within the Food and Drug Administration safety guideline of 2 degrees C. However, if a lead wire is broken, it is unsafe during MRI scans. Radiological examinations will be necessary to ensure the integrity of the implant.
引用
收藏
页码:367 / 373
页数:7
相关论文
共 19 条
[1]   HEATING EFFECTS OF METALLIC IMPLANTS BY MRI EXAMINATIONS [J].
BUCHLI, R ;
BOESIGER, P ;
MEIER, D .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (03) :255-261
[2]  
Chou CK, 1996, BIOELECTROMAGNETICS, V17, P195, DOI 10.1002/(SICI)1521-186X(1996)17:3<195::AID-BEM5>3.0.CO
[3]  
2-Z
[4]   ABSENCE OF RADIOFREQUENCY HEATING FROM AUDITORY IMPLANTS DURING MAGNETIC-RESONANCE-IMAGING [J].
CHOU, CK ;
MCDOUGALL, JA ;
CHAN, KW .
BIOELECTROMAGNETICS, 1995, 16 (05) :307-316
[5]  
CHOU CK, 1984, BIOELECTROMAGNETICS, V5, P435, DOI 10.1002/bem.2250050408
[6]   MAGNETIC-RESONANCE-IMAGING CAN CAUSE FOCAL HEATING IN A NONUNIFORM PHANTOM [J].
DAVIS, PL ;
SHANG, C ;
TALAGALA, L ;
PASCULLE, AW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (12) :1324-1327
[7]  
Durney C., 1986, Radiofrequency radiation dosimeter handbook, V4
[8]  
FAGAN LL, 1995, JMRI-J MAGN RESON IM, V5, P614
[9]   SPECIFIC ABSORPTION RATES OF ENERGY IN MAN MODELS EXPOSED TO CELLULAR UHF MOBILE-ANTENNA FIELDS [J].
GUY, AW ;
CHOU, CK .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (06) :671-680
[10]   SIMULATED BIOLOGICAL-MATERIALS FOR ELECTROMAGNETIC-RADIATION ABSORPTION STUDIES [J].
HARTSGROVE, G ;
KRASZEWSKI, A ;
SUROWIEC, A .
BIOELECTROMAGNETICS, 1987, 8 (01) :29-36